首页> 外文期刊>Journal of analytical chemistry >Effect of additives in a carrier gas on the behavior of ions in an ion-mobility increment spectrometer
【24h】

Effect of additives in a carrier gas on the behavior of ions in an ion-mobility increment spectrometer

机译:载气中添加剂对离子迁移率增量光谱仪中离子行为的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The effect of additives (ethanol, acetic acid, water, chloroform, and acetone) in air used as the carrier gas on the changes in the spectra of the H+(C6H5NH2) ion was studied for an ion-mobility increment spectrometer (IMIS) operating in tandem with a mass spectrometer (MS). It was shown that the presence of additives could improve both the resolution of the instrument and the sensitivity of the method. The resolving power was increased fourfold using air saturated with an ethanol vapor, while the sensitivity was increased 11-fold upon adding a mixture of ethanol and acetic acid to the carrier gas. The dependences on the concentration of the additives indicate that the formation of the clusters is the reason for these effects. The presence of an ethanol vapor in the carrier gas almost completely eliminates the effect of the present water vapor; therefore, air saturated with ethanol vapors can be used in the IMIS-MS instruments without drying.
机译:在离子迁移率增量光谱仪(IMIS)上,研究了用作载气的空气中的添加剂(乙醇,乙酸,水,氯仿和丙酮)对H +(C6H5NH2)离子光谱变化的影响。与质谱仪(MS)串联。结果表明,添加剂的存在可以提高仪器的分辨率和方法的灵敏度。使用充满乙醇蒸气的空气,分辨能力提高了四倍,而在载气中加入乙醇和乙酸的混合物后,灵敏度提高了11倍。对添加剂浓度的依赖性表明,团簇的形成是这些作用的原因。载气中乙醇蒸气的存在几乎完全消除了现有水蒸气的影响。因此,可以在IMIS-MS仪器中使用充满乙醇蒸气的空气而无需干燥。