首页> 外文期刊>Journal of Aerosol Science >Measurement of aerosol optical extinction at 532 nm with pulsed cavity ring down spectroscopy
【24h】

Measurement of aerosol optical extinction at 532 nm with pulsed cavity ring down spectroscopy

机译:脉冲腔衰荡光谱法测量532 nm处的气溶胶消光

获取原文
获取原文并翻译 | 示例
           

摘要

Accurate measurement of optical properties of aerosols is crucial for quantifying the influence of aerosols on climate.This article describes laboratory testing of a cavity ring down(CRD)system for measurement of the optical extinction of aerosol.The system is tested with well-characterized laboratory generated aerosols,and the measured extinctions agree within 5% of predictions by Mie theory.Extension of the lab prototype to field measurements of atmospheric aerosol is discussed.It is shown that the precision of measurements of the optical extinction of aerosol in the field will be determined by statistical fluctuations in the number of particles in the sample volume.Averaging times on the order of minutes are required to reduce statistical fluctuations to less than 5%.Improvements to the CRD system to reduce the averaging times are proposed.
机译:准确测量气溶胶的光学性质对于量化气溶胶对气候的影响至关重要。本文介绍了用于测量气溶胶光学消光的腔环下降(CRD)系统的实验室测试,并在性能良好的实验室进行了测试产生的气溶胶,并且测得的消光符合米氏理论预测值的5%。讨论了实验室原型扩展到大气气溶胶的现场测量。结果表明,在现场测量气溶胶的光学消光的精度将达到由样品体积中颗粒数量的统计波动确定。平均时间需要几分钟才能将统计波动降低到5%以下。建议对CRD系统进行改进以减少平均时间。