Effects of dual-energy subtraction chest radiography on detection of small pulmonary nodules with varying attenuation: receiver operating characteristic analysis using a phantom study.

机译:双能减影胸片对衰减变化不大的小肺结节的影响:使用体模研究的接收器工作特性分析。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: The aim of this study was to investigate the detectability of simulated pulmonary nodules with different X-ray attenuation by flat-panel detector (FPD) chest radiography using a dual-exposure dual-energy subtraction (DES) technique. MATERIALS AND METHODS: Using a FPD radiography system, we obtained 108 sets of chest radiographs of a chest phantom. They consisted of 54 sets each of chest radiographs with and without simulated nodules. Each data set contained a standard and a corresponding dual-energy subtracted chest radiograph (DES image). The diameters of the simulated nodules were 8, 10, and 12 mm, respectively; nodules of each size manifested attenuation of -450, -200, and 30 Hounsfield units (HU). We performed receiver operating characteristic (ROC) analysis to compare the observers' performance in detecting nodules. RESULTS: For the -450 HU nodules the mean areas under the ROC curve (AUC) without and with DES images were 0.66 and 0.77, respectively; the difference was significant (paired t-test, P < 0.01). For nodules with -200 and 30 HU, there was no significant difference in the AUC value (0.79 vs. 0.77, P = 0.13; 0.92 vs. 0.94, P = 0.17, respectively). CONCLUSION: The addition of DES images to standard chest radiographs improved the performance of radiologists charged with detecting simulated nodules with an attenuation of -450 HU.
机译:目的:本研究的目的是使用双曝光双能减影(DES)技术,通过平板探测器(FPD)胸部放射线照相术来研究具有不同X射线衰减的模拟肺结节的可检测性。材料与方法:使用FPD射线照相系统,我们获得了108套胸部幻影的胸部X光照片。他们由54套胸部X光片组成,有或没有模拟结节。每个数据集均包含标准品和相应的双能减胸片(DES图像)。模拟结节的直径分别为8、10和12 mm。每个大小的结节均表现出-450,-200和30的Hounsfield单位(HU)衰减。我们进行了接收器工作特性(ROC)分析,以比较观察者在检测结节中的表现。结果:对于-450 HU结节,在有和没有DES图像的情况下,ROC曲线下的平均面积分别为0.66和0.77。差异显着(配对t检验,P <0.01)。对于具有-200和30 HU的结节,AUC值无显着差异(分别为0.79对0.77,P = 0.13; 0.92对0.94,P = 0.17)。结论:在标准胸部X光片上增加了DES图像,改善了放射科医师检测模拟结节的能力,衰减了-450 HU。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号