We have implemented a simple, inexpensive, and fast procedure for validation and verification of the performance of pipettes mounted on automated liquid handlers (ALHs) as necessary for laboratories accredited under ISO 17025. A six- or seven-step serial dilution of OrangeG was prepared in quadruplicates in a flat-bottom 96-well microtiter plate, manually using calibrated pipettes. Each pipette of the liquid handler (1 - 8) dispensed a selected volume (1 - 200 μL) of OrangeG eight times into the wells of the microtiter plate. All wells contained a total of 200 μL liquid. The absorbance was read, and the dispensed volume of each pipette was calculated based on a plot of volume and absorbance of a known set of OrangeG dilutions. Finally, the percent inaccuracy (%d) and the imprecision (%CV) of each pipette were calculated. Using predefined acceptance criteria, each pipette was then either approved or failed. Failed pipettes were either repaired or the volume deviation was compensated for by applying a calibration curve in the liquid-handler software. We have implemented the procedure on a Sias Xantus, an MWGt TheONYX, four Tecan Freedom EVO, a Biomek NX Span-8, and four Biomek 3000 robots, and the methods are freely available. In conclusion, we have set up a simple, inexpensive, and fast solution for the continuous validation of ALHs used for accredited work according to the ISO 17025 standard. The method is easy to use for aqueous solutions but requires a spectrophotometer that can read microtiter plates.
机译:根据ISO 17025认可的实验室的需要,我们已经实施了一种简单,廉价且快速的程序来验证和验证安装在自动液体处理机(ALH)上的移液器的性能。准备了OrangeG的六步或七步连续稀释方法在平底96孔微量滴定板中一式四份,使用校准的移液器手动进行。每个移液器(1-8)的移液器都会将选定体积(1-200μL)的OrangeG八次分配到微量滴定板的孔中。所有孔中总共含有200μL液体。读取吸光度,并基于一组已知的OrangeG稀释液的吸光度和吸光度图计算每个移液器的分配体积。最后,计算每个移液器的不准确度百分比(%d)和不精确度(%CV)。然后,使用预定义的验收标准,每个移液器要么被批准要么失败。维修失败的移液器,或通过在液体处理机软件中应用校准曲线来补偿体积偏差。我们已经在Sias Xantus,MWGt TheONYX,四个Tecan Freedom EVO,一个Biomek NX Span-8和四个Biomek 3000机器人上实施了该程序,这些方法可免费获得。总之,我们已经建立了一个简单,便宜且快速的解决方案,用于根据ISO 17025标准对用于认可工作的ALH进行连续验证。该方法易于用于水溶液,但需要能够读取微量滴定板的分光光度计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号