Radiographic detection of artificially created horizontal root fracture using different cone beam CT units with small fields of view

机译:使用不同视场的锥形束CT单元对人工水平根部骨折进行X射线照相检测

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: To compare limited cone beam computerized tomography (CBCT) units with different field of views (FOVs) and voxel sizes in detecting artificially created horizontal root fracture (HRF) in extracted human teeth. Methods: Artificial HRF was created in the horizontal plane in 40 teeth. Another 40 intact teeth served as a control group. 80 teeth were placed in the respective maxillary anterior sockets of a human dry skull in groups. Six image sets were obtained: (1) Accuitomo 170,40 × 40mm FOV (0.080mm3); (2) Accuitomo 170,60 × 60mm FOV (0.125mm3); (3) Kodak 9000, 50 × 37 mm FOV (0.076 mm3); (4) Kodak 9000, 50 × 37mm FOV (0.100 mm3); (5) Vatech Pax-Duo3D 50 × 50mm FOV (0.080mm 3) and (6) Vatech Pax-Duo3D 85 × 85mm FOV (0.120 mm 3). Images were evaluated twice by five observers. Kappa values were calculated for observer agreement. Areas under the receiver operating characteristic (ROC) curves (Az values) were calculated, and the Az values for each image type were compared using t-tests (α = 0.05). Results: Intraobserver kappa coefficients ranged from 0.81 to 0.95 for the Accuitomo 170 images, from 0.80 to 0.92 for the Kodak 9000 images and from 0.76 to 0.95 for Vatech PanXDuo3D. The Az values for different image types and observers ranged from 0.93 to 0.97 for Accuitomo 170 images, from 0.93 to 0.98 for Kodak 9000 images and from 0.93 to 0.97 for the Vatech PanX-Duo3D images. No statistically significant differences (p 0.05) were found between the Az values. Conclusions: Limited CBCT units performed similarly in detecting simulated HRF.
机译:目的:比较具有不同视场(FOV)和体素大小的有限锥束计算机断层扫描(CBCT)单元,以检测提取的人牙中的人工创建的水平根部骨折(HRF)。方法:在40颗牙齿的水平面上创建人工HRF。另外40颗完整的牙齿作为对照组。将80颗牙齿成组放置在人类干颅骨的相应上颌前牙槽中。获得了六个图像集:(1)Accuitomo 170,40×40mm FOV(0.080mm3); (2)Accuitomo 170,60×60mm FOV(0.125mm3); (3)柯达9000,50×37毫米FOV(0.076毫米3); (4)柯达9000,50×37mm FOV(0.100 mm3); (5)Vatech Pax-Duo3D 50×50mm FOV(0.080mm 3)和(6)Vatech Pax-Duo3D 85×85mm FOV(0.120 mm 3)。五个观察者对图像进行了两次评估。计算卡伯值以达成观察员协议。计算接收器工作特性(ROC)曲线下的面积(Az值),并使用t检验(α= 0.05)比较每种图像类型的Az值。结果:对于Accuitomo 170图像,观察者内kappa系数的范围为0.81至0.95,对于Kodak 9000图像,其观察者kappa系数的范围为0.80至0.92,对于Vatech PanXDuo3D,该系数为0.76至0.95。对于Accuitomo 170图像,不同图像类型和观察者的Az值范围从0.93到0.97,对于Kodak 9000图像从0.93到0.98,对于Vatech PanX-Duo3D图像从0.93到0.97。在Az值之间没有发现统计学上的显着差异(p> 0.05)。结论:有限的CBCT单元在检测模拟HRF中的表现相似。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号