The purpose of this work is to study computationally the possibility of the application of a hybrid active noise control technique for magnetic resonance imaging (MRI) acoustic noise reduction. A hybrid control system combined with both feedforward and feedback loops embedded is proposed for potential application on active MRI noise reduction. A set of computational simulation studies were performed. Sets of MRI acoustic noise emissions measured at the patient's left ear location were recorded and used in the simulation study. By comparing three different control systems, namely, the feedback, the feedforward and the hybrid control, our results revealed that the hybrid control system is the most effective. The hybrid control system achieved approximately a 20-dB reduction at the principal frequency component. We concluded that the proposed hybrid active control scheme could have a potential application for MRI scanner noise reduction. (c) 2008 Elsevier Inc. All rights reserved.
机译:这项工作的目的是通过计算研究将混合有源噪声控制技术应用于磁共振成像(MRI)声噪声降低的可能性。提出了一种结合了前馈和反馈回路的混合控制系统,有望在主动MRI降噪中得到潜在应用。进行了一组计算仿真研究。记录在患者左耳位置测量的MRI声波发射集,并将其用于模拟研究。通过比较三种不同的控制系统,即反馈,前馈和混合控制,我们的结果表明混合控制系统是最有效的。混合控制系统在主频率分量处实现了大约20 dB的降低。我们得出的结论是,提出的混合主动控制方案可以在MRI扫描仪降噪方面有潜在的应用。 (c)2008 Elsevier Inc.保留所有权利。