首页> 外文期刊>Human brain mapping >Quantification of fMRI artifact reduction by a novel plaster cast head holder.
【24h】

Quantification of fMRI artifact reduction by a novel plaster cast head holder.

机译:通过新型石膏铸模头固定器对fMRI伪影减少的量化。

获取原文
获取原文并翻译 | 示例

摘要

In light of artifact-induced high variability of activation in fMRI repeat studies, we developed and tested a clinically useful plaster cast head holder (PCH) with improved immobilization, repositioning, and comfort. With PCH, there were considerably lower levels of translational and rotational head motion components compared to head fixation with conventional restraining straps (CRS). Rotational components cannot be fully compensated by realignment and lead to "false activations." In addition, task-correlated head motion, which highly increases the risk of artifacts, was considerably reduced with PCH, especially in a motion prone subject. Compared with PCH, head motion was 133% larger with CRS in a highly cooperative subject. With a motion prone subject, head motion range was increased by 769% (PCH: 0.9 mm, CRS: 7.8 mm), which may indicate the usefulness of PCH for restless patients. In functional activation maps, PCH alone yielded fewer residual motion artifacts than CRS + image registration. Subject tolerance of the head holder during the long measurement times of up to 2.5 hr was good, and slice orientation on different days confirmed the quality of repositioning.
机译:鉴于在fMRI重复研究中由假象引起的激活的高度可变性,我们开发并测试了具有改善的固定性,重新定位和舒适性的临床上有用的石膏石膏头保持器(PCH)。与使用传统约束带(CRS)固定头部相比,使用PCH时,平移和旋转头部运动分量的水平要低得多。旋转组件无法通过重新对齐完全补偿,并导致“错误激活”。另外,与任务相关的头部运动大大增加了伪影的风险,使用PCH可以大大降低,特别是在容易运动的受试者中。与PCH相比,在高度合作的受试者中,CRS的头部运动较大133%。对于容易运动的受试者,头部运动范围增加了769%(PCH:0.9毫米,CRS:7.8毫米),这可能表明PCH对躁动患者有用。在功能激活图中,与CRS +图像配准相比,仅PCH产生的残留运动伪像更少。在长达2.5小时的长时间测量中,头部支架的受检者耐受性良好,并且在不同日期的切片方向证实了重新定位的质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号