Numerical Analysis of Organ Doses Delivered During Computed Tomography Examinations Using Japanese Adult Phantoms with the WAZA-ARI Dosimetry System

机译:使用日本成人幻影与WAZA-ARI剂量测定系统对计算机断层摄影检查期间交付的器官剂量进行数值分析

获取原文
获取原文并翻译 | 示例

摘要

A dosimetry system for computed tomography (CT) examinations, named WAZA-ARI, is being developed to accurately assess radiation doses to patients in Japan. For dose calculations in WAZA-ARI, organ doses were numerically analyzed using average adult Japanese male (JM) and female (JF) phantoms with the Particle and Heavy Ion Transport code System (PHITS). Experimental studies clarified the photon energy distribution of emitted photons and dose profiles on the table for some multi-detector row CT (MDCT) devices. Numerical analyses using a source model in PHITS could specifically take into account emissions of x rays from the tube to the table with attenuation of photons through a beam-shaping filter for each MDCT device based on the experiment results. The source model was validated by measuring the CT dose index (CTDI). Numerical analyses with PHITS revealed a concordance of organ doses with body sizes of the JM and JF phantoms. The organ doses in the JM phantoms were compared with data obtained using previously developed systems. In addition, the dose calculations in WAZA-ARI were verified with previously reported results by realistic NUBAS phantoms and radiation dose measurement using a physical Japanese model (THRA1 phantom). The results imply that numerical analyses using the Japanese phantoms and specified source models can give reasonable estimates of dose for MDCT devices for typical Japanese adults.
机译:正在开发一种用于计算机断层扫描(CT)检查的剂量测定系统,名为WAZA-ARI,以准确评估对日本患者的辐射剂量。为了在WAZA-ARI中进行剂量计算,使用带有粒子和重离子迁移编码系统(PHITS)的日本成年人(JM)和女性(JF)体模对器官剂量进行了数值分析。实验研究阐明了某些多探测器行CT(MDCT)装置在工作台上发出的光子的光子能量分布和剂量分布。在PHITS中使用源模型进行的数值分析可以根据实验结果,具体考虑从射线管到工作台的x射线发射,以及通过每个MDCT设备的光束整形滤波器对光子的衰减。通过测量CT剂量指数(CTDI)验证了源模型。用PHITS进行的数值分析表明,器官剂量与JM和JF体模的体型一致。将JM体模中的器官剂量与使用先前开发的系统获得的数据进行比较。此外,通过现实的NUBAS体模和使用物理日本模型(THRA1体模)进行的辐射剂量测量,通过先前报道的结果验证了WAZA-ARI中的剂量计算。结果表明,使用日本人体模型和指定的源模型进行的数值分析可以为典型的日本成年人提供合理的MDCT装置剂量估计值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号