Status of Radiation Protection of Medical X-Ray Facilities in Greater Accra Region, Ghana

机译:加纳大阿克拉地区医疗X射线设施的辐射防护现状

获取原文
获取原文并翻译 | 示例

摘要

The status of radiation protection and safety of diagnostic medical x-ray facilities in the Greater Accra region in Ghana have been evaluated. In all, 62 medical facilities with 86 diagnostic x-ray units were considered for the survey. Out of the 86 diagnostic units, there were 56 general radiograph, 13 dental radiography, 9 fluoroscopy, 5 computed tomography, and 3 mammography machines. The parameters evaluated include the tube voltage, type of film processor, and the required protective measures in an x-ray department. It was observed that none of the protective measures or equipment were fully present in the diagnostic units except lead aprons. The radiation protection and safety measures in the medical facilities need to be strengthened to protect patients, staff, and the general public.
机译:已经评估了加纳大阿克拉地区的辐射防护状况和医用X射线诊断设备的安全性。总共考虑了具有86个诊断X射线装置的62个医疗机构。在86个诊断单位中,有56部普通射线照相,13部牙科X射线照相,9台透视检查,5台计算机断层摄影和3台乳腺X光摄影机。评估的参数包括管电压,胶片冲洗机的类型以及X射线部门所需的保护措施。观察到,除铅皮圈外,诊断单元中没有任何防护措施或设备完全存在。需要加强医疗机构的辐射防护和安全措施,以保护患者,医护人员和普通大众。