Effective penumbra and scalloping effect: a dosimetric study in multifield radiotherapy with multileaf collimator for prostate cancer treatment.

机译:有效的半影和扇贝效应:在多场放射治疗中使用多叶准直仪进行前列腺癌治疗的剂量学研究。

获取原文
获取原文并翻译 | 示例
           

摘要

AIMS AND BACKGROUND: The aim of the study was to test whether the multileaf collimator can be used to replace divergent alloy blocks for field shaping and to evaluate the dosimetric impact of the scalloping effect and the effective penumbra in multifield isocentric configurations routinely used at our department to treat early stage prostate tumor in supine patients. Deviations between measurements and treatment planning system calculations are also discussed in order to estimate the accuracy of effective penumbra calculations carried out by the software in the presence of blocks or a multileaf collimator. METHODS: The multileaf collimator installed on a dual energy (6 and 18 MV) linear accelerator Varian CLINAC 2100 C/D is an add-on component positioned below the standard jaws, with 40 computer-controlled opposed pairs of tungsten leaves. Transmission, effective penumbra and scalloping measurements were performed with films placed at different depths in a (30 x 30 x 20 cm3) acrylic phantom. A laser scanning photodensitometer was used to obtain the optical density and the relative dose profile. Effective penumbra and scalloping effect measured data were tested on a software phantom; the phantom, automatically performed with the treatment planning system, was a regular parallelepiped measuring 30 x 30 x 20 cm3 and the acrylic electronic density value. RESULTS: For one multileaf collimator-shaped field, the width of effective penumbra was about 2 mm wider than penumbra for cerrobend blocks, at a 45 degrees angle between leaf motion direction and the field edge. Collimator rotation, automatically performed by the treatment planning system, by minimizing the sum of over- and underblocked areas, reduces the differences between the multileaf collimator and blocks. The differences between measured and treatment planning system calculated data were within the treatment planning system dose calculation accuracy limits, as recommended in ICRU Report No. 42. CONCLUSIONS: Penumbra and, for the multileaf collimator, scalloping effect values seem to depend on the gantry angle, i.e., on the fields path of entry. The values for even fields are higher than the those for odd fields, because the dose gradient at the target edges is steeper for a single field than for two opposite fields, and the interplay of doses from the individual beams increases the distance between the isodose levels. Therefore, in order to reduce scalloping and effective penumbra values, it would be better to rotate the collimators, above all in even-number field techniques. In particular, the six-field technique, used mostly in prostate treatment, shows the same effective penumbra values with the multileaf collimator, with proper collimator rotation, and in blocked fields.
机译:目的和背景:本研究的目的是检验多叶准直仪是否可以代替发散的合金块进行场成形,并评估扇形效应和有效半影在本部门常规使用的多场等中心配置中的剂量学影响。治疗仰卧患者的早期前列腺肿瘤。还讨论了测量值与治疗计划系统计算之间的差异,以估计在存在块或多叶准直仪的情况下由软件执行的有效半影计算的准确性。方法:安装在双能(6和18 MV)线性加速器Varian CLINAC 2100 C / D上的多叶准直仪是位于标准钳口下方的附加组件,具有40对计算机控制的相对的钨叶对。透射,有效半影和扇贝测量是在(30 x 30 x 20 cm3)丙烯酸幻象中以不同深度放置的薄膜进行的。使用激光扫描光密度计获得光密度和相对剂量分布。有效半影和扇贝效应测量数据在软件模型上进行了测试;幻影是由治疗计划系统自动执行的,是规则的平行六面体,尺寸为30 x 30 x 20 cm3,具有丙烯酸电子密度值。结果:对于一个多叶准直器形的场,在叶片运动方向与场边缘之间成45度角时,有效半影的宽度比神经固定块的半影宽约2 mm。由治疗计划系统自动执行的准直仪旋转,通过最大程度地减少了上,下阻塞区域的总和,减少了多叶准直仪和模块之间的差异。根据ICRU报告第42号的建议,测量和治疗计划系统计算数据之间的差异在治疗计划系统剂量计算精度范围内。结论:半影以及对于多叶准直仪,扇贝效应值似乎取决于机架角度,即字段的输入路径。偶数场的值高于奇数场的值,因为单个场的目标边缘处的剂量梯度比两个相对场的陡峭,并且来自各个光束的剂量之间的相互作用增加了等剂量水平之间的距离。因此,为了减少扇贝和有效的半影值,最好首先将偶数场技术中的准直仪旋转。特别是,六场技术(主要用于前列腺治疗)在多叶准直仪,适当的准直仪旋转以及在遮挡场中显示出相同的有效半影值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号