CT image fusion as a tool for measuring in 3D the setup errors during conformal radiotherapy for prostate cancer.

机译:CT图像融合是一种用于在3D中测量前列腺癌保形放射治疗过程中设置错误的工具。

获取原文
获取原文并翻译 | 示例
           

摘要

AIMS AND BACKGROUND: The importance of optimal daily patient positioning has been stressed in order to ensure treatment reproducibility and gain in accuracy and precision. We report our data on the 3D setup uncertainty during radiation therapy for prostate cancer using the CT image fusion technique. METHODS: Ten consecutive patients scheduled for radiation therapy for prostate cancer underwent 5 prone position CT scans using an individualized immobilization cast. These different setups were analyzed using the image fusion module of the ERGO 3D-Line Medical System (Milan, Italy) treatment planning system. The isocenter and the body marker displacements were measured. RESULTS: The 3D isocenter dislocations were quantified: systematic error was sigma(3D) = 3.9 mm, whereas random error was sigma(3D) = 1 mm. The mean of the minimum displacements was 0.2 +/- 1 mm showing that the immobilization device used allows an accurate setup to be obtained. Single direction errors were also measured showing systematic errors, sigma(AP), = 2.6 mm, sigma(LL) = 0.6 mm, SigmaSI = 3 mm in the anterior-posterior, latero-lateral, superior-inferior direction, respectively. Related random errors were sigma(AP), = 1 mm, sigma(LL) = 0.6 mm, sigma(SI) = 1.2 mm. In terms of accuracy, our uncertainties are similar to those reported in the literature. CONCLUSIONS: By applying the CT image fusion technique, a 3D study on setup accuracy was performed. We demonstrated that the use of an individualized immobilization system for prostate treatment is adequate to obtain good setup accuracy, as long as a high-quality positioning control method, such as the stereoscopic X-ray-based positioning system, is used.
机译:目的和背景:为了确保治疗的可重复性和准确性和准确性的提高,已经强调了最佳的每日患者定位的重要性。我们使用CT图像融合技术报告我们在前列腺癌放射治疗期间3D设置不确定性的数据。方法:十名计划接受前列腺癌放射治疗的连续患者,使用个体化固定石膏进行了5次俯卧位CT扫描。使用ERGO 3D-Line医疗系统(意大利米兰)治疗计划系统的图像融合模块分析了这些不同的设置。测量了等角点和身体标记的位移。结果:量化了3D等角点错位:系统误差为sigma(3D)= 3.9 mm,而随机误差为sigma(3D)= 1 mm。最小位移的平均值为0.2 +/- 1 mm,这表明所使用的固定装置可实现精确的设置。还测量了单向误差,分别显示了前后方向,外侧水平方向,上下方向的系统误差sigma(AP)= 2.6 mm,sigma(LL)= 0.6 mm,SigmaSI = 3 mm。相关的随机误差为sigma(AP)= 1 mm,sigma(LL)= 0.6 mm,sigma(SI)= 1.2 mm。在准确性方面,我们的不确定性与文献报道的相似。结论:通过应用CT图像融合技术,对安装精度进行了3D研究。我们证明,只要使用高质量的定位控制方法(例如基于X射线的立体定位系统),就可以使用个性化固定系统进行前列腺治疗足以获得良好的设置精度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号