首页> 外文期刊>Transfusion: The Journal of the American Association of Blood Banks >Recommendations for optimized settings of the Amicus Crescendo cell separator for the collection of CD34+ progenitor cells.
【24h】

Recommendations for optimized settings of the Amicus Crescendo cell separator for the collection of CD34+ progenitor cells.

机译:为收集CD34 +祖细胞而优化Amicus Crescendo细胞分离器设置的建议。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

BACKGROUND: CD34+ PBPCs for autologous transplantation purposes are collected by leukapheresis procedures on automated cell separators. In this study, the influence of different parameters on collection efficiency (CE) of the Amicus Crescendo cell separator (Baxter) was investigated. STUDY DESIGN AND METHODS: A total of 146 PBPC collections with Amicus cell separators were performed in 56 patients with either settings recommended by the manufacturer or modified settings to identify variables that have a significant and important impact on CE. RESULTS: By use of a standard setting with a cycle volume of 1400 mL, CE significantly decreases when patients' preapheresis peripheral blood WBC counts are between 25,000 and 35,000 per micro L. CE can be improved if cycle volume is reduced to 1000 mL. If WBC concentrations exceed 55,000 per micro L before apheresis, CE also significantly decreases despite of reduced cycle volume. Additionally, high flow rates greater than 60 mL per minute significantly reduce CE. CONCLUSION: Parameters influencing the outcome of CD34+ PBPC collections were identified, such as patients' WBC count, cycle volume, and whole blood flow rate. An optimized adjustment of these variables will further increase the CE of the device.
机译:背景:用于白细胞移植的CD34 + PBPC通过白细胞分离术收集在自动细胞分离器上。在这项研究中,研究了不同参数对Amicus Crescendo细胞分离器(Baxter)的收集效率(CE)的影响。研究设计和方法:在56例患者中进行了146例带有Amicus细胞分离器的PBPC收集,这些患者采用制造商建议的设置或修改后的设置,以识别对CE有重要影响的变量。结果:通过使用标准设置,循环量为1400 mL,当患者术前血浆WBC计数在每微升25,000至35,000之间时,CE显着降低。如果循环量减少至1000 mL,则可以改善CE。如果在单采前WBC浓度超过55,000 / micro L,尽管循环量减少,CE也会显着降低。此外,每分钟大于60 mL的高流速会显着降低CE。结论:确定了影响CD34 + PBPC收集结果的参数,例如患者的WBC计数,周期量和全血流量。这些变量的优化调整将进一步提高设备的CE。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号