Relationship Between Cyclosporine Concentrations Obtained Using the Roche Cobas Integra and Abbott TDx Monoclonal Immunoassays in Pre-dose and Two Hour Post-dose Blood Samples from Kidney Transplant Recipients.

机译:在肾脏移植受者的给药前和给药后两小时血样中,使用Roche Cobas积分和Abbott TDx单克隆免疫测定获得的环孢菌素浓度之间的关系。

获取原文
获取原文并翻译 | 示例
           

摘要

Current evidence suggests that cyclosporine (CsA) concentration in blood samples taken 2 hours after Neoral microemulsion (Novartis Pharmaceuticals; East Hanover, NJ) administration (C ) predicts clinical events in transplant patients better than the pre-dose (trough) concentration (C ). Similarly, previous findings have shown that the metabolites/CsA ratio is substantially lower in C than in C samples; however the between-monoclonal immunoassay differences for C samples have received little attention in the literature. In 56 C samples and 60 C samples from renal transplant patients, CsA levels were determined using the monoclonal fluorescence polarization immunoassay (mFPIA) from Abbott (Abbott Park, IL) and the homogeneous enzyme immunoassay technique (HEIT) from Roche Diagnostics (Basel, Switzerland). In both cases a high correlation coefficient between the results was obtained (r >/= 0.971), with a linear regression for C samples: mFPIA = 1.47 HEIT + 22.0 and for C samples: mFPIA = 1.11 HEIT + 71.96. The difference between the linear regression slopes was statistically significant (P < 0.001), and the mFPIA/HEIT ratio was significantly higher for C than for C samples (P < 0.001).(2) (0) (2) (0) (2) (0) (2) (0) (2) (0) (2)
机译:目前的证据表明,在使用神经微乳剂(Novartis Pharmaceuticals; East Hanover,NJ)2小时后采集的血液样本中,环孢素(CsA)浓度预测的移植患者临床事件要好于给药前(谷)浓度。 。同样,以前的发现表明,C中的代谢物/ CsA比值明显低于C样品。但是,C样品的单克隆之间免疫测定差异在文献中很少受到关注。在来自肾移植患者的56 C和60 C样品中,使用来自Abbott(伊利诺伊州阿伯特帕克)的单克隆荧光偏振免疫测定(mFPIA)和来自Roche Diagnostics(瑞士巴塞尔)的均相酶免疫测定技术(HEIT)测定CsA水平。 )。在这两种情况下,都获得了结果之间的高相关系数(r> / = 0.971),C样本的线性回归:mFPIA = 1.47 HEIT + 22.0,C样本:mFPIA = 1.11 HEIT + 71.96。线性回归斜率之间的差异具有统计学意义(P <0.001),C的mFPIA / HEIT比显着高于C样品(P <0.001)。(2)(0)(2)(0)( 2)(0)(2)(0)(2)(0)(2)

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号