首页> 外文期刊>The Journal of Urology >Efficacy, safety and patient satisfaction outcomes of the AMS 700CX inflatable penile prosthesis: results of a long-term multicenter study. AMS 700CX Study Group.
【24h】

Efficacy, safety and patient satisfaction outcomes of the AMS 700CX inflatable penile prosthesis: results of a long-term multicenter study. AMS 700CX Study Group.

机译:AMS 700CX充气式阴茎假体的功效,安全性和患者满意度:一项长期的多中心研究结果。 AMS 700CX研究小组。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: We performed a long-term multicenter study of the AMS 700CX 3-piece inflatable penile prosthesis, focusing on longevity, morbidity and patient satisfaction in men implanted up to 134 months with a median followup of 47.7 months. MATERIALS AND METHODS: We performed a large scale retrospective multicenter study in 2 phases. Phase 1 was a medical record review of 372 men who underwent implantation with the AMS 700CX penile prosthesis from 1987 to 1996 by 7 frequent penile prosthesis implanters. Phase 2 included a structured telephone interview of 207 patients by a neutral observer. RESULTS: For the 372 men in phase 1 mean device mechanical reliability plus or minus standard deviation was 92.1% + or - 3.3% after 3 and 86.2% + or - 4.6% after 5 years. Patient age was 21 to 79 years (mean 57.6 + or - 11.0) at implantation. The etiology of erectile dysfunction was vascular in 27.7% of the cases, Peyronie's disease in 16.9%, diabetes mellitus in 12.9% and radical surgery in 11.6%. Of the men 55.6% received previous treatment for erectile dysfunction. Postoperative infection and device malfunction developed in 3.2% and 17.5% of the cases, respectively. Of the 207 men interviewed in phase 2, 86% still had an AMS 700CX penile prosthesis implanted, including 87.1% with erection suitable for coitus. Currently 79% of those with a device use it at least twice monthly and 88.2% would recommend an implant to a relative or friend. CONCLUSIONS: The AMS 700CX penile implant produced suitable erection and excellent patient satisfaction at long-term followup in the majority of men. Implant reliability is excellent and postoperative morbidity is low.
机译:目的:我们对AMS 700CX 3片可充气阴茎假体进行了长期的多中心研究,重点研究了植入134个月的男性的寿命,发病率和患者满意度,平均随访47.7个月。材料与方法:我们分两个阶段进行了大规模的回顾性多中心研究。第一阶段是一项医疗记录,回顾了1987年至1996年期间由7位频繁的阴茎假体植入者植入AMS 700CX阴茎假体的372名男性。第二阶段包括由中立的观察员对207位患者进行的结构化电话采访。结果:对于372名处于第1阶段的男性,其平均设备机械可靠性的正负标准偏差为3年后为92.1%+或-3.3%,五年后为86.2%+或-4.6%。植入时患者年龄为21至79岁(平均57.6 +或-11.0)。勃起功能障碍的病因学是血管的27.7%,佩罗尼氏病16.9%,糖尿病12.9%,根治性手术11.6%。在这些男性中,有55.6%曾接受过勃起功能障碍的治疗。术后感染和器械故障分别占3.2%和17.5%。在第2阶段接受采访的207名男性中,仍有86%植入了AMS 700CX阴茎假体,其中87.1%的勃起适合于性交。目前,有设备的人中有79%每月至少使用两次,有88.2%的人会向亲戚或朋友推荐植入物。结论:在大多数男性的长期随访中,AMS 700CX阴茎植入物产生了适当的勃起效果,并获得了出色的患者满意度。植入物可靠性极佳,术后发病率低。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号