Methyl Methacrylate Levels in Orthopedic Surgery: Comparison of Two Conventional Vacuum Mixing Systems

机译:矫形外科手术中甲基丙烯酸甲酯的水平:两种常规真空混合系统的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Poly-methyl methacrylate bone cements contain methyl methacrylate (MMA), which is known for its sensitizing and toxic properties. Therefore, in most European countries and in the USA, guidelines or regulations exist for occupational exposures. The use of vacuum mixing systems can significantly reduce airborne MMA concentrations during bone setting. Our goal was to test two commonly used vacuum mixing systems (Palamix and Optivac) using Palacos R bone cement for their effectiveness at preventing MMA vapor release in a series of standardized trials in a laboratory as well as in an operating theatre. MMA was quantified every second over a period of 3min using a photoionization detector (MiniRAE 3000) device positioned in the breathing area of the user. Significant differences in MMA mean vapor concentrations over 180 s were observed in the two experimental spaces, with the highest mean concentrations (7.61 and 7.98 ppm for Palamix and Optivac, respectively) observed in a laboratory with nine air changes per hour and the lowest average concentrations (1.06 and 1.12 ppm for Palamix and Optivac, respectively) in an operating theatre with laminar flow ventilation and 22 air changes per hour. No significant differences in overall MMA concentrations were found between the two vacuum mixing systems in either location. Though, differences were found between both systems during single mixing phases. Thus, typical handling of MMA in orthopedic procedures must be seen as not harmful as concentrations do not reach the short-term exposure limit of 100 ppm. Additionally, laminar airflow seems to have an influence on lowering MMA concentrations in operation theatres.
机译:聚甲基丙烯酸甲酯骨水泥包含甲基丙烯酸甲酯(MMA),该物质以其敏化和有毒特性而闻名。因此,在大多数欧洲国家和美国,存在有关职业暴露的准则或规定。在骨骼固定过程中,使用真空混合系统可以显着降低空气中的MMA浓度。我们的目标是在实验室和手术室进行的一系列标准化试验中,使用Palacos R骨水泥测试两种常用的真空混合系统(Palamix和Optivac)在防止MMA蒸气释放方面的有效性。使用位于用户呼吸区域中的光电离检测器(MiniRAE 3000),在3分钟内每秒对MMA进行定量。在两个实验空间中,观察到超过180 s的MMA平均蒸汽浓度存在显着差异,在实验室中每小时观察到9次换气,平均平均浓度最低,平均浓度最高(Palamix和Optivac分别为7.61和7.98 ppm) (在具有层流通风和每小时22次换气的手术室中(对于Palamix和Optivac,分别为1.06和1.12 ppm)。在两个位置的两个真空混合系统之间,总MMA浓度均无显着差异。但是,在单个混合阶段,两个系统之间发现了差异。因此,必须将整形外科手术中的MMA常规处理视为无害的,因为其浓度不能达到100 ppm的短期暴露极限。另外,层流气流似乎对降低手术室中的MMA浓度有影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号