Stability measurements of craniofacial implants by means of resonance frequency analysis. A clinical pilot study.

机译:通过共振频率分析测量颅面植入物的稳定性。临床试验研究。

获取原文
获取原文并翻译 | 示例

摘要

Nineteen patients previously treated with 52 implants for anchorage of craniofacial prostheses were subjected to implant stability measurements by means of resonance frequency analysis (RFA), six months to 15 years after implant placement. The resonance frequency (RF) of a transducer attached to the implant abutment was measured by using a frequency response analyser, a personal computer (PC) and dedicated software. Statistically significant higher RF values were seen for implants in the temporal bone as compared to implants in the nose and periorbital regions. There was a positive correlation with time since implant placement for the period from six months up to seven years. It was concluded that the preliminary results suggest that implant stability increases with time and that implants in temporal bone are more stable than implants in the bone in the nose and periorbital regions, probably reflecting differences in bone density.
机译:在植入植入后6个月至15年内,通过共振频率分析(RFA)对19例先前用52枚植入物进行颅面假体固定的患者进行了植入物稳定性测量。通过使用频率响应分析仪,个人计算机(PC)和专用软件来测量连接到种植体基台的换能器的共振频率(RF)。与鼻子和眶周区域的植入物相比,颞骨的植入物在统计学上具有较高的RF值。自植入植入物以来的六个月至七年时间与时间呈正相关。结论是,初步结果表明,植入物的稳定性随时间增加,并且颞骨中的植入物比鼻子和眶周区域中的骨中的植入物更稳定,这可能反映了骨密度的差异。