Whole blood metal ion measurement reproducibility between different laboratories

机译:不同实验室之间的全血金属离子测量重现性

获取原文
获取原文并翻译 | 示例
       

摘要

Monitoring patients' metal ion blood concentrations can be useful in cases of problematic metal on metal hip implants. Our objective was to evaluate the reproducibility of metal ion level values measured by two different laboratories. Whole blood samples were collected in 46 patients with metal on metal hip arthroplasty. For each patients, two whole blood samples were collected and analyzed by two laboratories. Laboratory 1 had higher results than laboratory 2. There was a clinically significant absolute difference between the two laboratories, above the predetermined threshold, 35% of Cr samples and 38% of Co samples. All laboratories do not use the same technologies for their measurements. Therefore, decision to revise a metal on metal hip arthroplasty should rely on metal ion trends and have to be done in the same laboratory. (C) 2014 Elsevier Inc. All rights reserved.
机译:监测患者的金属离子血液浓度在金属髋关节植入物上有问题的金属的情况下很有用。我们的目标是评估由两个不同实验室测得的金属离子水平值的可重复性。在46例金属髋关节置换术中收集了全血样本。对于每个患者,收集了两个全血样本,并由两个实验室进行了分析。实验室1的结果高于实验室2。两个实验室之间在临床上存在绝对的临床差异,高于预定阈值,35%的Cr样品和38%的Co样品。所有实验室都不使用相同的技术进行测量。因此,对金属髋关节置换术中的金属进行修改的决定应取决于金属离子的趋势,并且必须在同一实验室中进行。 (C)2014 Elsevier Inc.保留所有权利。