首页> 外文期刊>The Journal of arthroplasty >Intermediate-Term Results of 142 Single-Design, Rotating-Hinge Implants: Frequent Complications May Not Preclude Salvage of Severely Affected Knees
【24h】

Intermediate-Term Results of 142 Single-Design, Rotating-Hinge Implants: Frequent Complications May Not Preclude Salvage of Severely Affected Knees

机译:142例单设计旋转铰接植入物的中期结果:频繁的并发症可能无法挽救严重受累的膝盖

获取原文
获取原文并翻译 | 示例
           

摘要

Rotating-hinge knee prostheses have low survivorship and high complications except in primary arthroplasties in elderly patients. We retrospectively reviewed 142 single third-generation design, rotating hinge prostheses (11 primary procedures and 131 revisions) at 57 months follow up. Implant survival was 73 %. Successful two-stage reimplantation for prosthetic infection was 78.4% but new infection rate was 22%. The tibial component was durable while the femoral component was problematic. We observed only one patellar maltracking and no polyethylene wear. A third generation rotating-hinge arthroplasty reconstruction was reliable in complex problems. Outcomes in primary situations were excellent. Complications were the rule rather than the exception in revisions. With timely intervention, attention to soft tissue coverage, and realistic expectations, complications were contained and functional benefits were appreciable. (C) 2015 Elsevier Inc. All rights reserved.
机译:旋转铰接式膝关节假体的存活率低,并发症多,但老年患者的初次关节置换术除外。我们回顾了57个月的随访情况,回顾了142个第三代单一设计,旋转铰链假体(11个主要步骤和131个修订版本)。植入物存活率为73%。成功的两阶段假体感染再植入成功率为78.4%,但新感染率为22%。胫骨组件坚固耐用,而股骨组件有问题。我们仅观察到一种pa骨错位,没有聚乙烯磨损。第三代旋转铰链置换术重建在复杂问题上是可靠的。在主要情况下的结果非常好。复杂是规则,而不是修订中的例外。通过及时的干预,对软组织覆盖的关注以及切合实际的期望,可以控制并发症的发生,并获得明显的功能益处。 (C)2015 Elsevier Inc.保留所有权利。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号