Effect of Taper Design on Trunnionosis in Metal on Polyethylene Total Hip Arthroplasty

机译:锥度设计对聚乙烯全髋关节置换术中金属耳轴的影响

获取原文
获取原文并翻译 | 示例
           

摘要

This study examines how taper design affects corrosion and fretting at the head trunnion surface. All hip prostheses retrieved between 1999 and 2013 with 28 mm/+0 heads were selected, resulting in 44 cobalt-chrome-on-polyethylene implants, representing six taper designs. Mean implantation time: 8.9 +/- 3.7 years. The femoral head tapers were scored for fretting and corrosion using the Goldberg scale as both a combined score and by three zones (apex, central and base). There was no difference in age (P = 0.34), BMI (P = 0.29), or implantation time (P - 0.19) between taper groups. The 11/13 taper had the highest combined corrosion and fretting score, but no difference (P = 0.22) between groups for combined scores (P = 0.22 for corrosion, P = 0.19 for fretting). In a zone-specific analysis, the 11/13 taper had highest corrosion score at base zone (P = 0.02). Taper design had a significant effect on corrosion at base of trunnion. Crown Copyright (C) 2015 Published by Elsevier Inc. All rights reserved.
机译:这项研究研究了锥度设计如何影响头部耳轴表面的腐蚀和微动。选择了1999年至2013年之间所有头部为28 mm / + 0的髋关节假体,共制作了44种聚乙烯铬钴合金植入物,代表了六种锥度设计。平均植入时间:8.9 +/- 3.7年。使用戈德堡量表作为综合评分和三个区域(顶点,中央和底部)对股骨头锥度进行微动和腐蚀评分。锥度组之间的年龄(P = 0.34),BMI(P = 0.29)或植入时间(P-0.19)没有差异。 11/13锥度的腐蚀和微动磨损综合得分最高,但各组之间的复合分数没有差异(P = 0.22)(腐蚀的P = 0.22,微动的P = 0.19)。在特定区域分析中,11/13锥度在基部区域具有最高的腐蚀评分(P = 0.02)。锥度设计对耳轴基部的腐蚀有显着影响。 Crown版权(C)2015,由Elsevier Inc.出版。保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号