Vitamin E diffusion into highly cross-linked polyethylene (E-XLPE) is a method for enhancing oxidative stability of acetabular liners. The purpose of this study was to evaluate in vivo penetration of E-XLPE using radiostereometric analysis (RSA). Eighty-four hips were recruited into a prospective 10-year RSA. This is the first evaluation of the multicenter cohort after 3-years. All patients received E-XLPE liners (E1, Biomet) and porous-titanium coated cups (Regenerex, Biomet). There was no difference (P = 0.450) in median femoral head penetration into the E-XLPE liners at 3-years comparing cobalt-chrome heads (-0.028 mm; inter-quartile range (IQR) -0.065 to 0.047) with ceramic heads (-0.043 mm, IQR -0.143 to 0.042). The 3-year follow-up indicates minimal E-XLPE liner penetration regardless of head material and minimal early cup movement. (C) 2015 Elsevier Inc. All rights reserved.
机译:维生素E扩散到高度交联的聚乙烯(E-XLPE)中是增强髋臼衬板氧化稳定性的一种方法。这项研究的目的是使用放射立体分析(RSA)评估E-XLPE在体内的渗透率。募集了84个髋关节纳入预期的10年RSA。这是三年后对多中心队列的首次评估。所有患者均接受E-XLPE衬垫(E1,Biomet)和多孔钛涂层杯(Regenerex,Biomet)。与钴铬陶瓷头(-0.028 mm;四分位间距(IQR)-0.065至0.047)相比,陶瓷铬头(-0.028 mm;四分位间距(IQR)-0.065至0.047)在3年时,股骨头向E-XLPE内衬的穿透率无差异(P = 0.450) -0.043毫米,IQR -0.143至0.042)。三年的随访表明,无论头部材料如何,E-XLPE衬套的穿透力都最小,并且杯的早期移动极小。 (C)2015 Elsevier Inc.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号