摘要

Long term total hip arthroplasty follow-up has demonstrated bearing surface wear and failure of fixation as the major modes of failure. The purpose of this study was to evaluate the minimum 10 year results of a third generation cementless acetabular component using moderately cross-linked polyethylene liners. 150 primary THAs were evaluated clinically for need for revision and activity evaluated using accelerometers, WOMAC, Tegner and UCLA activity-level scores. Radiographs were evaluated for wear, osteolysis and loosening. We compared this cohort to three of our previously reported cohorts. The only revision was unrelated to liner wear and all acetabular components were bone ingrown. There were no cases of acetabular osteolysis and one case of femoral osteolysis. Average steps per year were 1.59 million. Mean steady state wear rate averaged 0.04 mm/year. At 10-years, excellent fixation and low wear were demonstrated using a cementless acetabular construct and moderately cross-linked polyethylene liner.
机译:长期的全髋关节置换术后随访表明,轴承表面磨损和固定失败是主要的失败方式。这项研究的目的是评估使用中度交联聚乙烯内衬的第三代非骨水泥髋臼组件的至少10年结果。临床上评估了150个原发性THA的修订需求,并使用加速度计,WOMAC,Tegner和UCLA活动水平评分对活动进行了评估。对射线照相进行磨损,溶骨和松动评估。我们将该队列与先前报告的三个队列进行了比较。唯一的改变与内衬磨损无关,所有髋臼组件均向内生长。没有髋臼骨溶解病例和股骨骨溶解病例。每年平均步数为159万。平均稳态磨损率平均为0.04毫米/年。使用非骨水泥髋臼结构和中度交联的聚乙烯内衬,在使用10年时,表现出出色的固定性和低磨损。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号