首页> 外文期刊>The Journal of arthroplasty >The TRAC PS mobile-bearing prosthesis: design rationale and in vivo 3-dimensional laxity.
【24h】

The TRAC PS mobile-bearing prosthesis: design rationale and in vivo 3-dimensional laxity.

机译:TRAC PS移动假体:设计原理和体内3维松弛。

获取原文
获取原文并翻译 | 示例
       

摘要

We present a posterior stabilized mobile-bearing prosthesis, TRAC PS, which has congruent contact from full extension to full flexion, allows for freedom of internal-external rotation, and has an automatic posterior shift in tibiofemoral contact on the tibia to maximize the quadriceps lever arm in flexion. TRAC PS has 2 radii of curvature in the sagittal plane, 1 for the distal femoral condyles and 1 for the posterior femoral condyles, as does the normal knee. The distal and posterior femoral condyles articulate congruently in the inner tracks or the outer tracks of the polyethylene bearing, respectively. Anterior or posterior sliding of the femoral condyles on the bearing or of the bearing on the tibial tray cannot occur, providing inherent anterior and posterior stability. Three-dimensional knee laxity testing was performed on 17 patients from 12 months to 25 months after total knee arthroplasty with the TRAC PS and on 18 healthy control subjects of similar ages. Normal ligament balancing and normal internal-external rotational laxity were achieved with the TRAC PS prosthesis. Anterior and posterior laxity in the patients with TRAC PS was significantly reduced compared with that of the control subjects.
机译:我们提出了一种后稳定的活动关节假体TRAC PS,它具有从完全伸展到完全屈曲的全接触,允许内部-外部旋转的自由度,并且胫骨上的股骨股骨接触自动向后移动,以最大化股四头肌杠杆屈臂。 TRAC PS的矢状面曲率半径为2,股骨远端con为1,股骨后dy为1,正常膝关节也是如此。股骨远端和后con分别在聚乙烯轴承的内轨道或外轨道内完全铰接。股骨con在支座上或胫骨托上的支座不会向前或向后滑动,从而提供固有的前后稳定性。在TRAC PS进行全膝关节置换术后12个月至25个月期间,对17位患者进行了三维膝关节松弛测试,并对18位年龄相似的健康对照者进行了三维测试。 TRAC PS假体可实现正常的韧带平衡和正常的内外旋转松弛度。与对照组相比,TRAC PS患者的前后松弛度明显降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号