首页> 外文期刊>The Journal of arthroplasty >Fatigue Fracture in Dual Modular Revision Total Hip Arthroplasty Stems. Failure Analysis and Computed Tomography Diagnostics in Two Cases
【24h】

Fatigue Fracture in Dual Modular Revision Total Hip Arthroplasty Stems. Failure Analysis and Computed Tomography Diagnostics in Two Cases

机译:双模块修订全髋关节置换术茎中的疲劳骨折。两种情况下的故障分析和计算机断层扫描诊断

获取原文
获取原文并翻译 | 示例
           

摘要

We report on two patients with fracture of a modular, tapered and distally fixed, uncemented titanium revision hip stem, not previously described. A failure analysis revealed that the cause of the fractures was the development of fatigue cracks in the mid-stem cobalt-chromium modular junction ending in corrosion-fatigue failure. No material defects or stress risers were found in any of the implants. The diameter of the mid-stem modular junction might be undersized for use in heavy and active patients. We also report a new way of detecting an undisplaced fracture at the modular junction, using the scout image from a computed tomography (CT) scan; a technique that can be used when plain radiographs are inconclusive.
机译:我们报道了两名患者,先前没有描述过模块化,锥形和远端固定,非骨水泥钛修订版髋关节干骨折。失效分析表明,断裂的原因是中部钴-铬模块连接处疲劳裂纹的发展,最终导致腐蚀疲劳失效。在任何植入物中均未发现材料缺陷或应力升高。中柄模块化连接的直径可能过小,无法用于重度和活跃患者。我们还报告了一种使用计算机断层扫描(CT)扫描中的侦察图像检测模块接合处未移位骨折的新方法。当普通X射线照片不确定时可以使用的一种技术。