Reliability of a New Method for Evaluating Femoral Stem Positioning After Total Hip Arthroplasty Based on Stereoradiographic 3D Reconstruction

机译:基于立体放射3D重建的评估全髋关节置换术后股骨干定位的新方法的可靠性

获取原文
获取原文并翻译 | 示例
       

摘要

The goal of this study was to validate a new method for determining femoral stem positioning based on 3D models derived from the EOS biplanar system. Independents observers measured stem anteversion and femoral offset using CT scan and EOS system of 28 femoral stems implanted in composite femurs. In parallel, the same parameters were measured on biplanar lower limb radiographs acquired from 30 patients who had undergone total hip arthroplasty. CT scanner and biplanar X-ray measurements on composite femurs were highly correlated: 0.94 for femoral offset (P < 0.01), 0.98 for stem anteversion (P<0.01). The inter and intra-observer reproducibility when measuring composite bones was excellent with both imaging modalities as when measuring femoral stem positioning in patients with the biplanar X-ray system. (C) 2014 Elsevier Inc. All rights reserved.
机译:这项研究的目的是基于从EOS双平面系统衍生的3D模型,验证一种确定股骨柄定位的新方法。独立观察者使用CT扫描和EOS系统测量了植入复合股骨的28个股骨干的茎前倾和股骨偏移。同时,在从30例行全髋关节置换术的患者获得的双平面下肢X光片上测量了相同的参数。 CT扫描和双平面X线片对股骨的高度相关性:股骨偏移0.94(P <0.01),茎前倾0.98(P <0.01)。与在双平面X射线系统患者中测量股骨柄位置时一样,在测量复合骨时,观察者之间和观察者之间的可重复性在两种成像方式下均表现出色。 (C)2014 Elsevier Inc.保留所有权利。