The present study evaluated the effects of extent of conformity of post-cam design on contact area and stress at post-cam mechanism using 4 different posterior-stabilized prostheses. TRAC and Alpina with full-conformed post-cams exhibited the largest contact area at 90 degrees and 120 degrees . PFC sigma RPF with partial conformed post-cam had the largest contact area at 150 degrees . Scorpio NRG with less conformed post-cam had smaller contact area than the others. Lifting of femoral component decreased contact area and increased contact stress of TRAC and Alpina. Recent modifications of post-cam design have increased contact area, contributing to lower contact stress. None of these prostheses exhibited constant low contact stress throughout flexion. Further modifications of post-cam mechanism are necessary to provide lower contact stress throughout deep knee flexion.
机译:本研究使用4种不同的后稳定假体评估了凸轮设计的一致性程度对凸轮结构后接触面积和应力的影响。 TRAC和Alpina具有完整的后置凸轮,在90度和120度处显示最大的接触面积。具有部分贴合后置凸轮的PFC sigma RPF在150度处具有最大的接触面积。 Scorpio NRG的后置凸轮轮廓较差,接触面积较小。股骨组件抬起减少了接触面积,并增加了TRAC和Alpina的接触应力。凸轮后设计的最新修改增加了接触面积,有助于降低接触应力。这些假肢在整个屈曲过程中均未表现出恒定的低接触应力。凸轮后机构的进一步修改是必要的,以在整个膝盖深度弯曲中提供较低的接触应力。