首页> 外文期刊>The Journal of arthroplasty >Late dissociation of an alumina-on-alumina bearing modular acetabular component.
【24h】

Late dissociation of an alumina-on-alumina bearing modular acetabular component.

机译:氧化铝在氧化铝上承载的模块化髋臼组件的后期解离。

获取原文
获取原文并翻译 | 示例
           

摘要

Alumina-on-alumina bearings for total hip arthroplasty have been introduced as a promising alternative to reduce wear debris and to increase the life expectancy of the prosthesis. We report a case of a late dissociation of an alumina-on-alumina bearing modular acetabular component, which occurred 2 years, 8 months after surgery. Detailed analysis of the retrieved prosthesis suggested that the cause of the failure may be strong rotational torque developed by a roughening of the bearing's alumina surface caused by edge loading (microseparation). The strong friction torque in articulation may explain the higher reported rate of mechanical loosening of the acetabular component in alumina-on-alumina bearing total hip prostheses.
机译:已经引入了用于全髋关节置换术的氧化铝对氧化铝轴承,作为减少磨损碎片并增加假体寿命的一种有希望的替代方法。我们报告了一个案例,即术后2年8个月发生了氧化铝在氧化铝上的模块化髋臼组件的晚期分离。对取回的假体的详细分析表明,失效的原因可能是由于边缘载荷(微分离)导致轴承的氧化铝表面变粗糙而产生的强大旋转扭矩。关节中的强摩擦扭矩可能解释了承载氧化铝的全髋关节假体中髋臼组件机械松动的报道率较高。