【24h】

Iron(III) ionophores based on formylsalicylic acid derivatives as sensors for ion-selective electrodes

机译:基于甲酰水杨酸衍生物的铁(III)离子载体作为离子选择电极的传感器

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Novel iron(III)-selective PVC membrane electrodes based on formylsalicylic acid derivatives were studied. The electrode based on p-chloroaniline-3-formylsalicylic acid as a sensor, containing potassium tetrakis(4-chlorophenyl)borate as a lipophilic salt and o-nitrophenyl octyl ether as a plasticizer, gave the best performance. The electrode exhibits a good Nernstian response for 10(-1)-5.0 x 10(-5) mol l(-1) FeCl3 with a slope of 20 mV per decade. It shows a high selectivity for iron(III) in comparison with alkali, alkaline earth and heavy metal ions. The electrode response and selectivity remained almost unchanged for at least 1 month. The effects of plasticizers, membrane supports, lipophilic salts and pH on the potential response of the electrode were also studied. The electrode was successfully applied to the determination of iron(III) contents in some rocks. [References: 32]
机译:研究了基于甲酰水杨酸衍生物的新型铁(III)-选择性PVC膜电极。以对氯苯胺-3-甲酰基水杨酸为传感器的电极,以四(4-氯苯基)硼酸钾为亲脂盐,邻硝基苯基辛基醚为增塑剂,表现出最佳的性能。电极对10(-1)-5.0 x 10(-5)mol l(-1)FeCl3表现出良好的能斯特响应,每十年的斜率为20 mV。与碱金属,碱土金属和重金属离子相比,它对铁(III)具有很高的选择性。电极响应和选择性在至少1个月内几乎保持不变。还研究了增塑剂,膜载体,亲脂性盐和pH对电极电位响应的影响。该电极已成功应用于某些岩石中铁(III)含量的测定。 [参考:32]