Wear particles from metal-on-metal total hip replacements: effects of implant design and implantation time

机译:金属对金属的全髋关节置换术产生的磨损颗粒:植入物设计和植入时间的影响

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Detailed characterization of wear particles is necessary to understand better the implant wear mechanisms and the periprosthetic tissue response. The purposes of the present study were to compare particle characteristics of current with older designs of metal-on-metal (MM) total hip replacements (THRs), and to determine the effect of implantation time on wear particle characteristics. Metal wear particles isolated from periprosthetic tissues from 19 patients with MM THRs of current and older designs and at different implantation times (very short, longer, and very long) were studied using transmission electron microscopy and energy dispersive X-ray analysis. The particles from the current design implants with implantation times of not more than 15 months (very short term) were almost exclusively round to oval chromium oxide particles. In all other cases, although the predominance was still round to oval chromium oxide particles, greater proportions of cobalt-chromium-molybdenum (Co-Cr-Mo) particles, mainly needle-shaped, were detected. Very long term THRs implanted for more than 20 years had the highest percentage of needle-shaped Co-Cr-Mo particles. Particle lengths were not markedly different between the different designs and implantation times except for the current design implants of not more than 15 months, which had a significantly smaller mean length of 39 nm. In conclusion, the implant design did not seem to have a significant influence on particle characteristics whereas the implantation time appeared to have the most effect on the particles. It should be noted that, because of the limited number of tissue retrievals available, some uncertainty remains regarding the generality of these findings.
机译:磨损颗粒的详细表征对于更好地了解植入物的磨损机理和假体周围组织的响应是必要的。本研究的目的是将电流的颗粒特征与旧式金属对金属(MM)全髋关节置换(THR)设计进行比较,并确定植入时间对磨损颗粒特征的影响。使用透射电子显微镜和能量色散X射线分析技术研究了从19例当前和较旧设计的MM THR患者,不同植入时间(非常短,更长和非常长)的假体周围组织中分离出的金属磨损颗粒。当前设计植入物的粒子植入时间不超过15个月(非常短期),几乎都是圆形至椭圆形的氧化铬粒子。在所有其他情况下,尽管占主导地位的仍然是椭圆形的氧化铬颗粒,但可以检测到更大比例的钴-铬-钼(Co-Cr-Mo)颗粒,主要是针状。植入超过20年的非常长期的THR具有最高比例的针状Co-Cr-Mo颗粒。在不同设计和植入时间之间,粒子长度没有显着差异,但当前设计的植入物不超过15个月,其平均长度显着较小,仅为39 nm。总之,植入物的设计似乎对颗粒的特性没有显着影响,而植入时间似乎对颗粒的影响最大。应当指出的是,由于可用的组织取回数量有限,这些发现的普遍性仍存在一些不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号