首页> 外文期刊>Proceedings of the Institution of Mechanical Engineers, Part H. Journal of Engineering in Medicine >Comparative wear tests of ultra-high molecular weight polyethylene and cross-linked polyethylene
【24h】

Comparative wear tests of ultra-high molecular weight polyethylene and cross-linked polyethylene

机译:超高分子量聚乙烯和交联聚乙烯的对比磨损试验

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Wear particle-induced osteolysis is a major concern in hip implant failure. Therefore, recent research work has focussed on wear-resistant materials, one of the most important of which is cross-linked polyethylene. In view of this, the objective of this study was to compare the in vitro wear performance of cross-linked polyethylene to traditional ultra-high molecular weight polyethylene. In order to mimic appropriate in vivo conditions, a novel high-capacity wear tester called a circularly translating pin-on-disc was used. The results of this in vitro study demonstrated that the wear rate for crosslinked polyethylene was about 80% lower than that of conventional ultra-high molecular weight polyethylene. This difference closely matches in vivo results reported in the literature for total hip replacements that use the two biopolymers. The in vitro results were also verified against ASTM F732-00 (standard test method for wear testing of polymeric materials for use in total joint prostheses). The 50-station circularly translating pin-on-disc proved to be a reliable device for in vitro wear studies of orthopaedic biopolymers.
机译:磨损引起的骨溶解是髋关节植入物衰竭的主要问题。因此,最近的研究工作集中在耐磨材料上,其中最重要的一种是交联聚乙烯。有鉴于此,本研究的目的是比较交联聚乙烯与传统超高分子量聚乙烯的体外磨损性能。为了模拟合适的体内条件,使用了一种新型的大容量磨损测试仪,称为圆形平移针盘。这项体外研究的结果表明,交联聚乙烯的磨损率比常规超高分子量聚乙烯低约80%。这种差异与文献中报道的使用两种生物聚合物进行全髋关节置换的体内结果非常吻合。还根据ASTM F732-00(用于全关节假体的聚合物材料的磨损测试的标准测试方法)验证了体外结果。 50工位圆盘针盘式平移证明是骨科生物聚合物体外磨损研究的可靠设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号