Intra-operator and inter-operator reliability of manual and semiautomated measurement of fetal nuchal translucency: A cross sectional study

机译:人工和半自动测量胎儿颈部半透明性的操作者内和操作者间可靠性:一项横断面研究

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: The goal of this study was to examine the intra-operator and inter-operator differences of the manual and semiautomated nuchal translucency (NT) measurements and to evaluate if these differences alter women's risk status. Methods: A cross sectional study was performed. Two operators obtained manual and semiautomated NT measurements of 153 NT images. The maximal acceptable difference in NT measurements within and between operators was 0.15mm. Intra and inter-operator differences were analyzed by the paired Student's t-test and homogeneity of variances by the Levene's test. Intra-operator and inter-operator agreement were quantified with Bland and Altman's limits of agreement, and changes in women's risk status were tested with the binomial test. Results: Intra-operator agreement was high for each of the measurement methods. Operator 1 had lower SDS for manual measurements. Conversely, operator 2 had lower SDS of the differences for semiautomated measurements, although the SD never reached the same level as operator 1. Inter-operator agreement was highest for the semiautomated measurements. Changes in risk status occurred between the manual and inner-middle method resulting in different clinical policies in up to 1 out of 20 cases. Conclusion: Well-trained operators do not seem to benefit from the use of the semiautomated measurement methods.
机译:目的:本研究的目的是检查人工和半自动口腔半透明(NT)测量值在操作者内部和操作者之间的差异,并评估这些差异是否会改变女性的危险状况。方法:进行横断面研究。两名操作员获得了153个NT图像的手动和半自动NT测量值。运营商内部及其之间的NT测量值的最大可接受差值为0.15mm。通过配对的Student's t检验分析了操作员内部和操作员之间的差异,通过Levene检验分析了方差的均一性。运营商内部和运营商之间的协议通过Bland和Altman的协议限制进行了量化,并通过二项式检验对女性风险状况的变化进行了测试。结果:每种测量方法的操作员内部协议都很高。操作员1的SDS较低,用于手动测量。相反,尽管SD从未达到与操作员1相同的水平,但操作员2的半自动差值的SDS较低。手动方法和内部-中间方法之间发生了风险状态变化,导致多达20个病例中有1个发生了不同的临床策略。结论:训练有素的操作员似乎无法从半自动测量方法中受益。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号