Point-of-Care Measurements on a Neonatal Intensive Care Unit Using the OMNI-S Blood Gas Analyzer

机译:使用OMNI-S血气分析仪在新生儿重症监护病房进行即时护理测量

获取原文
获取原文并翻译 | 示例
           

摘要

Near-patient testing of blood sugar, bilirubin, and hematocrit (Hct) has been performed for many years on neonatal intensive care units. The aim of this study was to compare traditional measurements using Ascensia Elite XL Blood Glucometers, bilimeters, and Hct readers with the Roche OMNI-S analyzer. Simultaneous blood samples were collected for gas machine analysis (maximum, 200 muL) and centrifuged capillary samples (150 muL) for standard analysis. Between 80 and 168 measurements were each made of serum bilirubin (n = 113), Hct (n = 168), glucose (n = 100), sodium (n = 81), and potassium (n = 80) and compared on the OMNI-S analyzer. The correlation coefficients (r) were the following: bilirubin, 0.98; Hct, 0.85; glucose, 0.91; sodium, 0.83; potassium, 0.80. The slope of the regression lines were the following: bilirubin, 0.96 (95% confidence interval [CI], 0.92-0,99); Hct, 0.98 (95% CI, 0.90-1.07); glucose, 0.88 (95% CI, 0.82-0,95); sodium, 0.78 (95% CI, 0.67-0.94); potassium, 1.00 (95% CI, 0.92- 1.14). There was excellent agreement between the Roche OMNI-S analyzer and traditional measurement systems used on the unit; the system can be used to accurately measure bilirubin, Hct, and glucose and gives a good reflection of serum electrolytes in our patient population. The use of a single system simplifies staff training and equipment maintenance and reduces running costs and volumes of blood samples.
机译:新生儿重症监护病房已经对患者的血糖,胆红素和血细胞比容(Hct)进行了近年患者测试。这项研究的目的是比较使用Ascensia Elite XL血糖仪,胆管仪和Hct读数器与Roche OMNI-S分析仪进行的传统测量。收集血样同时进行气机分析(最大200μL)和离心毛细管样品(150μL)进行标准分析。在80到168次之间分别测量了血清胆红素(n = 113),Hct(n = 168),葡萄糖(n = 100),钠(n = 81)和钾(n = 80)并在OMNI上进行了比较-S分析仪。相关系数(r)如下:胆红素为0.98;胆红素为0.98。每小时0.85;葡萄糖0.91;钠0.83;钾0.80。回归线的斜率如下:胆红素0.96(95%置信区间[CI],0.92-0,99); Hct,0.98(95%CI,0.90-1.07);葡萄糖,0.88(95%CI,0.82-0,95);钠0.78(95%CI,0.67-0.94);钾,1.00(95%CI,0.92-1.14)。罗氏OMNI-S分析仪与设备上使用的传统测量系统之间达成了极好的协议。该系统可用于准确测量胆红素,Hct和葡萄糖,并能很好地反映我们患者群体中的血清电解质。单个系统的使用简化了人员培训和设备维护,并降低了运行成本和血液样本量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号