首页> 外文期刊>Water, Air, and Soil Pollution >Feasibility of Field Portable Near Infrared (MR) Spectroscopy to Determine Cyanide Concentrations in Soil
【24h】

Feasibility of Field Portable Near Infrared (MR) Spectroscopy to Determine Cyanide Concentrations in Soil

机译:现场便携式近红外(MR)光谱法测定土壤中氰化物浓度的可行性

获取原文
获取原文并翻译 | 示例
       

摘要

Vicinities of manufactured gas plants were often contaminated with solid iron-cyanide complexes as a result of the coal gasification process. During the remediation of affected soils, knowledge about contaminant concentrations is crucial, but laboratory methods are often expensive and time consuming. Rapid and non-destructive field methods for contaminant determination permit an analysis of large sample numbers and hence, facilitate identification of'hot spots' of contamination. Diffuse near infrared reflectance spectroscopy has proven to be a reliable analytical tool in soil investigation. In order to determine the feasibility of a Poly-chromix Handheld Field Portable Near-Infrared Analyzer (FP NIR), various sample preparation methods were examined, including homogenizing, sieving, drying, and grinding. Partial least squares calibration models were developed to determine near infrared (NIR) spectral responses to the cyanide concentration in the soil samples. As a control, the contaminant concentration was determined using conventional flow injection analysis. The experiments revealed that portable near-infrared spectrometers could be a reliable device for detecting cyanide concentrations >2,400 mgkg~(-1) in the field and > 1,750 mgkg~(-1) after sample preparation in the laboratory. We found that portable NIR spectrometry cannot replace traditional laboratory analyses due to high limits of detection, but that it could be used for identification of contamination 'hot spots'.
机译:煤气化过程的结果是,人造煤气厂附近经常被固态的氰化铁络合物污染。在修复受影响的土壤期间,有关污染物浓度的知识至关重要,但是实验室方法通常很昂贵且耗时。快速,无损的现场污染物测定方法可以分析大量样品,因此有助于识别污染物的“热点”。漫反射近红外反射光谱已被证明是土壤研究中可靠的分析工具。为了确定Poly-chromix手持式现场便携式近红外分析仪(FP NIR)的可行性,检查了各种样品制备方法,包括均质,筛分,干燥和研磨。开发了偏最小二乘校准模型,以确定对土壤样品中氰化物浓度的近红外(NIR)光谱响应。作为对照,使用常规流动注射分析法确定污染物浓度。实验表明,便携式近红外光谱仪可以作为一种可靠的设备,用于现场检测氰化物浓度> 2,400 mgkg〜(-1),实验室制备后检测氰化物浓度> 1,750 mgkg〜(-1)。我们发现便携式NIR光谱由于检测限高而不能替代传统的实验室分析,但可用于识别污染“热点”。

著录项

  • 来源
    《Water, Air, and Soil Pollution》 |2012年第8期|p.5495-5504|共10页
  • 作者单位

    Geopedology and Landscape Development,Brandenburg University of Technology,Konrad-Wachsmann-Allee 6,03046 Cottbus, Germany;

    Central Analytical Laboratory,Brandenburg University of Technology,Konrad-Wachsmann-Allee 6,03046 Cottbus, Germany;

    Soil Protection and Recultivation,Brandenburg University of Technology,Konrad-Wachsmann-Allee 6,03046 Cottbus, Germany;

    Geopedology and Landscape Development,Brandenburg University of Technology,Konrad-Wachsmann-Allee 6,03046 Cottbus, Germany;

    Soil Protection and Recultivation,Brandenburg University of Technology,Konrad-Wachsmann-Allee 6,03046 Cottbus, Germany;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    NIR spectroscopy; iron-cyanide complexes; partial least squares calibration; flow injection analysis;

    机译:近红外光谱氰化铁络合物;局部最小二乘校准;流动注射分析;
  • 入库时间 2022-08-17 13:40:42