Application of European Commission Reference Dose Levels in Some Common CT Examinations in King AbdulAziz University Hospital

机译:欧盟委员会参考剂量水平在阿卜杜勒·阿齐兹国王大学医院一些常见的CT检查中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

"CT doses now represent the greatest single source of medical exposure and the use of CT is increasing rapidly. In some departments, CT scanning is about 15 % of total examinations but may account for 70 % of the dose delivered". Many ways are found to describe and measure radiation dose in CT. Recently, European Guidelines (EG) on quality criteria for CT were published by the European Commission (EC), in which two dose descriptors, weighted computed tomography dose index (CTDI_W) and dose-length product (DLP), were proposed as reference dose levels (RDLs). It should not be exceeded without special clinical reasons. These dose levels can be used as a guideline if national reference dose levels have not yet been established. In this study measurements are carried out using a pencil shaped ionization chamber of active length 100 mm (RADCAL), and TLD 100 chips. The objective of this study is to investigate routine protocols of some common CT examinations utilized in the KAU Hospital in terms of imaging technique and radiation dose and to compare results with EC RDLs.
机译:“现在,CT剂量代表了最大的医疗暴露源,并且CT的使用正在迅速增加。在某些部门中,CT扫描约占总检查量的15%,但可能占所提供剂量的70%”。发现了许多描述和测量CT辐射剂量的方法。最近,欧洲委员会(EC)发布了有关CT质量标准的欧洲指南(EG),其中提出了两个剂量描述符,即加权计算机断层摄影剂量指数(CTDI_W)和剂量长度乘积(DLP)作为参考剂量级别(RDL)。没有特殊的临床原因,不应超过此范围。如果尚未确定国家参考剂量水平,则可以将这些剂量水平用作指导。在这项研究中,使用有效长度为100 mm的铅笔形电离室(RADCAL)和TLD 100芯片进行测量。这项研究的目的是就影像学技术和放射剂量研究在KAU医院中使用的一些常见CT检查的常规方案,并将结果与​​EC RDL进行比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号