Computer-aided technology can decrease the difficulty associated with arthroscopic surgery. Unlike a larger incision that exposes the whole joint, a small arthroscopic incision limits the surgeon to view the joint only from the arthroscope. Our computer-aided system for hip arthroscopy addresses this loss of joint visibility by (1) tracking tool position with a linkage of encoders, and (2) indicating the tool position relative to the patient anatomy. This paper presents a study of user performance to verify the effectiveness of this computer-aided system for hip arthroscopy. A user study was completed to determine if the computer-aided system could help reduce task completion time and tool path length. Ten participants completed a simple navigation task with and without the assistance of the computer-aided system. A time reduction of 38% and a 72% tool path-length decrease was achieved with the computer-aided system, confirming its effectiveness.
机译:计算机辅助技术可以减少与关节镜手术相关的困难。与暴露整个关节的较大切口不同,较小的关节镜切口会限制外科医生仅从关节镜观察关节。我们的用于髋关节镜检查的计算机辅助系统通过(1)通过编码器的链接跟踪工具位置,以及(2)指示工具相对于患者解剖结构的位置,解决了关节可见度的下降。本文提出了一项对用户表现的研究,以验证该计算机辅助系统用于髋关节镜检查的有效性。完成了一项用户研究,以确定计算机辅助系统是否可以帮助减少任务完成时间和工具路径长度。有或没有计算机辅助系统的帮助,十名参与者完成了一个简单的导航任务。使用计算机辅助系统可以节省38%的时间,并减少72%的刀具路径长度,从而证实了其有效性。