Novel tamoxifen-selective potentiometric electrodes based on various ion associations of tamoxifen were developed. The most suitable membrane composition, which has the best potentiometric characteristics, was examined under steady-state conditions. The membrane composed of (w/w) 3% tamoxifen-phosphomolibdate ion pair, 32% poly(vinylchloride), and 65% 2-nitrophenyloctylether was determined as the membrane with the best analytical properties. On the condition of , the fabricated electrode exhibited linear responses over the concentration range of M tamoxifen solutions () with a sensitivity of 42.9 ± 0.3 mV/decade, detection limit of M and short response time of s. The influences of some alkaline, alkaline earth metals, biologically significant molecules, and some pharmaceutical species over electrode responses were studied. The proposed electrode was applied for the determination of tamoxifen contents of some pharmaceutical formulations by using standard addition method. The potentiometric analysis results obtained for the pharmaceutical formulations were in good harmony with the results obtained by high performance liquid chromatography at 95% confidence level.
机译:基于他莫昔芬的各种离子缔合的新型他莫昔芬选择性电位电极被开发出来。在稳态条件下检查了具有最佳电位特性的最合适的膜组合物。确定由(w / w)3%的他莫昔芬-磷钼酸根离子对,32%的聚氯乙烯和65%的2-硝基苯基辛基醚组成的膜是具有最佳分析性能的膜。在的条件下,所制备的电极在M他莫昔芬溶液()的浓度范围内表现出线性响应,灵敏度为42.9±0.3 mV /十年,M的检测极限和s的短响应时间。研究了一些碱金属,碱土金属,生物学上重要的分子和某些药物种类对电极响应的影响。所建议的电极用于通过标准添加方法测定某些药物制剂中他莫昔芬的含量。药物制剂获得的电位分析结果与高效液相色谱法在95%置信水平下获得的结果相吻合。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号