首页> 外文期刊>Sensors and Actuators. B, Chemical >Use of a chromatographic column to improve the selectivity of the gas sensors: first approach towards a miniaturised device and selective with hydrogen fluoride vapours
【24h】

Use of a chromatographic column to improve the selectivity of the gas sensors: first approach towards a miniaturised device and selective with hydrogen fluoride vapours

机译:使用色谱柱改善气体传感器的选择性:迈向小型化设备的第一种方法,并采用氟化氢蒸气进行选择性处理

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Tin dioxide-based gas sensors were laid out at the exit of a chromatographic column for the selective detection of hydrogen fluoride in the atmosphere. After having selected a stationary phase adapted to this particular molecule, the sensor's signals were studied using various experimental parameters. In this paper a complete description of the experimental device and of the electrical responses is given. For the experiment, the gas sample was formed by a gas mixture of HF, O_2 and N_2. It was found that the optimal mode of detection corresponds to the sensitive element (SnO_2) having a temperature near to 500℃ with a carrier gas of oxidising nature (synthetic air). Moreover, it was found that with this experimental configuration, a minimum concentration of about 0.8 ppm was detectable.
机译:在色谱柱出口处布置了基于二氧化锡的气体传感器,用于选择性检测大气中的氟化氢。选择适合该特定分子的固定相后,使用各种实验参数研究传感器的信号。在本文中,给出了对实验装置和电响应的完整描述。对于实验,气体样品是由HF,O_2和N_2的气体混合物形成的。发现最佳的检测方式对应于具有氧化性质的载气(合成空气),温度接近500℃的敏感元素(SnO_2)。此外,发现利用该实验配置,可检测到约0.8ppm的最小浓度。