This work presents a comparative study of the 6 MeV percentage depth dose curves for photon beams of VARIAN Clinac 2300 C/D, ELEKTA Synergy Platform, and SIEMENS Primus Linacs. The measurements were performed for a set of 7 field sizes (5×5, 10×10, 15×15, 20×20, 25×25, 30×30, 40×40) cm~2, keepingthe same conditions (e.g., geometry, ion chamber voltage and polarity, incremental step and direction). During our study we used the Mephysto MC~2 software and two semiflex ion chambers from Physikalisch-Technische Werkstatten GmbH, calibrated in water under ~(60)Co gamma ray beam.
机译:这项工作对VARIAN Clinac 2300 C / D,ELEKTA Synergy Platform和SIEMENS Primus Linacs的光子束的6 MeV百分比深度剂量曲线进行了比较研究。在7个场大小(5×5、10×10、15×15、20×20、25×25、30×30、40×40)cm〜2的一组条件下进行测量,并保持相同的条件(例如,几何形状,离子室电压和极性,增量步长和方向)。在我们的研究中,我们使用了Mephysto MC〜2软件和Physikalisch-Technische Werkstatten GmbH的两个semiflex离子室,该室在〜(60)Coγ射线束下于水中校准。