首页> 外文期刊>Romanian reports in physics >A comparative study on 6 MeV photon beam percentage depth dose of VARIAN CLINAC 2300 c/d, ELEKTA synergy platform, and Siemens Primus LINACS
【24h】

A comparative study on 6 MeV photon beam percentage depth dose of VARIAN CLINAC 2300 c/d, ELEKTA synergy platform, and Siemens Primus LINACS

机译:VARIAN CLINAC 2300 c / d,ELEKTA协同平台和Siemens Primus LINACS的6 MeV光子束百分比深度剂量的比较研究

获取原文
           

摘要

This work presents a comparative study of the 6 MeV percentage depth dosecurves for photon beams of VARIAN Clinac 2300 C/D, ELEKTA Synergy Platform,and SIEMENS Primus Linacs. The measurements were performed for a set of 7 fieldsizes (55, 1010, 1515, 2020, 2525, 3030, 4040) cm2, keeping thesame conditions (e.g., geometry, ion chamber voltage and polarity, incremental step anddirection). During our study we used the Mephysto MC2 software and two semiflex ionchambers from Physikalisch-Technische Werkst tten GmbH, calibrated in water under60Co gamma ray beam.
机译:这项工作对VARIAN Clinac 2300 C / D,ELEKTA Synergy平台和SIEMENS Primus直线加速器的光子束的6 MeV百分比深度剂量曲线进行了比较研究。对一组7个场尺寸(55、1010、1515、2020、2525、3030、4040)cm2进行测量,并保持相同的条件(例如,几何形状,离子室电压和极性,增量阶跃和方向)。在我们的研究中,我们使用了Mephysto MC2软件和Physikalisch-Technische Werkst tten GmbH的两个Semiflex离子室,该离子室在60 Co伽玛射线束的水中进行了校准。