Robinson‐type nuclear quadrupole resonance spectrometer adapted to field‐effect transistors

机译:适用于场效应晶体管的Robinson型核四极共振光谱仪

获取原文
           

摘要

A nuclear quadrupole resonance spectrometer, the Robinson circuit adapted with field‐effect transistors, has operated satisfactorily up to 30 MHz with rf levels ranging from 40 mV to 2.6 V peak to peak across the sample coil. The S/N ratio of 14N in hexamethylenetetramin is 85 at room temperature and 150 at 77 K, with a sample volume of 6 cm3 and a time constant of 10 s. The spectrometer is also capable of measuring the spin‐lattice relaxation time by the saturation method.
机译:装有场效应晶体管的Robinson电路是一个核四极共振光谱仪,在高达30 MHz的频率下令人满意地工作,其rf电平在整个采样线圈之间达到40 mV至2.6 V的峰峰值。六亚甲基四胺中14 N的信噪比在室温下为85,在77 K下为150,样品量为6 cm3,时间常数为10 s。光谱仪还能够通过饱和法测量自旋晶格弛豫时间。