We present the design and operation of a test cassette for exposure of samples to radiation environments at the National Ignition Facility. The cassette provides options for square and round samples and exposure areas; the cassette provides for multiple levels of filtration on a single sample, which allows dynamic range in experiments. The samples had normal lines of sight to the x-ray source in order to have uniform x-ray illumination. The incident x-radiation onto the samples was determined by the choice of filter thicknesses and materials. The samples were held at precise locations, accurate to within a few hundred microns, in the target chamber in order to have a known fluence incident. In the cassette, the samples were held in place in such a way that a minimal “line contact” allows them to have the maximal mechanical response to the x-ray load. We present postshot images of the debris found on films used for filters, and pre- and postexposure specimens. © 2010 American Institute of Physics Article Outline INTRODUCTION REQUIREMENTS AND DESIGN CONSIDERATIONS Sample holders and sample contact Filter pairs and collimator design Sample alignment Cassette alignment OPERATION AND RESULTS Filter film debris analysis Sample exposures SUMMARY AND CONCLUSIONS
机译:我们介绍了在国家点火设施中将样品暴露于辐射环境中的测试盒的设计和操作。暗盒提供方形和圆形样品以及曝光区域的选件;样品盒可在单个样品上提供多种过滤水平,从而允许动态范围的实验。样品具有与X射线源垂直的视线,以便获得均匀的X射线照度。通过选择滤光片的厚度和材料来确定入射到样品上的X射线。将样品保持在目标腔室中精确的位置,精确到几百微米以内,以产生已知的注量入射。在样品盒中,样品被固定在适当的位置,以使最小的“线接触”使样品对X射线负荷具有最大的机械响应。我们展示了用于过滤器的胶片以及曝光前和曝光后样本上发现的碎片的快照后图像。 ©2010美国物理研究所文章大纲简介要求和设计注意事项样品架和样品触点过滤器对和准直仪设计样品对准盒式对准操作和结果滤膜碎屑分析样品暴露总结和结论

著录项

  • 来源
    《Review of Scientific Instruments 》 |2010年第7期| p.1-9| 共9页
  • 作者单位

    Lawrence Livermore National Laboratory, Livermore, California 94551, USA;

    Lawrence Livermore National Laboratory, Livermore, California 94551, USA;

    Lawrence Livermore National Laboratory, Livermore, California 94551, USA;

    Lawrence Livermore National Laboratory, Livermore, California 94551, USA;

    Lawrence Livermore National Laboratory, Livermore, California 94551, USA;

    Gray Research, Inc., 655 Discovery Drive, Suite 300, Huntsville, Alabama 35806, USA;

    Gray Research, Inc., 655 Discovery Drive, Suite 300, Huntsville, Alabama 35806, USA;

    Gray Research, Inc., 655 Discovery Drive, Suite 300, Huntsville, Alabama 35806, USA;

    Naval Research Laboratory, Washington, DC 20375, USA;

    Naval Research Laboratory, Washington, DC 20375, USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    ignition; X-rays;

    机译:点火;X射线;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号