首页> 外文期刊>Radiation Protection Dosimetry >Monte carlo calculated effective dose to Teenage girls from computed tomography Examinations
【24h】

Monte carlo calculated effective dose to Teenage girls from computed tomography Examinations

机译:蒙特卡洛从计算机断层扫描检查中计算出对少女的有效剂量

获取原文
获取原文并翻译 | 示例
           

摘要

Effective doses from CT to paediatric patients are not common in the literature. This article reports some effective doses to teenage girls from CT examinations. The voxel computational model ADELAIDE, representative of a 14-year-old girl, was scaled in size by ±5/100 to represent also 11-12-year-old and 16-year-old girls. The EGS4 Monte Carlo code was used to calculate the effective dose from chest, abdomen and whole torso CT examinations tot he three versions of ADELAIDE using a 12 kV spectrum. For the whole torso CT examination, in order of increasing model size, the effective doses were 9.0, 8.2 and 7.8 mSv per 100 mA.s. Data are presented that allow the estimation of effective dose from CT examinations of the torso for girls between the ages of 11 and 16.
机译:从CT到儿科患者的有效剂量在文献中并不普遍。本文从CT检查中报告了对少女的一些有效剂量。代表14岁女孩的体素计算模型ADELAIDE的大小按±5/100缩放,以代表11-12岁和16岁的女孩。使用EGS4蒙特卡洛代码,使用12 kV频谱从三个版本的ADELAIDE的胸部,腹部和整个躯干CT检查中计算有效剂量。对于整个躯干CT检查,按照模型尺寸增大的顺序,有效剂量为每100 mA.s 9.0、8.2和7.8 mSv。提供的数据可以通过11岁至16岁女孩的躯干CT检查估算出有效剂量。