Personal neutron dosemeters from seven US Department of Energy (DOE) laboratories were mailed to the European Laboratory for Particle Physics (CERN) and irradiated using the well-characterised CERN reference radiation facility (CERF). Neutron dose equivalents determined using the DOE personal dosemeters have been compared to the reference dose equivalent as determined using a tissue-equivalent proportional counter (TEPC). In the 0.5 to 5 mSv dose equivalent range, the comparison of results suggests that the neutron personal dosemeters in use at DOE facilities are capable of estimating dose equivalents for high energy neutrons to within a factor of at least 2 or 3. If a field-specific calibration factor is used to correct the dose equivalent responses, the agreement with the reference dose equivalent for these dosemeters can be improved to better than about 25 to 65. Suggestions for future intercomparison studies and methods of improving personal dosimetry at accelerator facilities are discussed.
机译:来自美国能源部(DOE)七个实验室的个人中子剂量仪已邮寄到欧洲粒子物理实验室(CERN),并使用功能齐全的CERN参考辐射设施(CERF)进行辐射。使用DOE个人剂量仪确定的中子剂量当量已与使用组织当量比例计数器(TEPC)确定的参考剂量当量进行了比较。在0.5至5 mSv剂量当量范围内,结果比较表明,DOE设施中使用的中子个人剂量仪能够将高能中子的剂量当量估计为至少2或3倍。使用特定的校准因子来校正剂量当量响应,这些剂量计与参考剂量当量的一致性可以改善到优于约25到65。讨论了未来的比较研究和改善加速器设施个人剂量的方法的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号