The purpose of this study was to develop an equation with which to determine the tube current to be used in order to obtain a certain image noise level for differently sized children undergoing multi-slice computed tomography examination. The relationship between image noise and detector dose for different examination protocols was established for a LightSpeed Ultra, an eight slice CT from GEMS, using homogeneous water phantoms of different sizes. Three different anatomical areas (head, thorax and abdomen) were studied in 111 patients between 0 and 17 y of age. The mean ratio between the calculated and the measured noise in patient images was established for the different areas. Head examinations showed the best correlation (measured-to-calculated noise ratio = 1.01). In the thorax, the calculated noise was generally higher than the measured noise (ratio = 0.74), and in the abdomen, the opposite result was found (ratio = 1.20).
机译:这项研究的目的是开发一个方程式,用于确定要使用的管电流,以便为进行多层计算机断层扫描检查的不同大小的孩子获得一定的图像噪声水平。使用不同尺寸的均质水体模,为LightSpeed Ultra(GEMS的八层CT)建立了不同检查方案下图像噪声与检测器剂量之间的关系。在111位0至17岁的患者中研究了三个不同的解剖区域(头部,胸部和腹部)。针对不同区域建立了患者图像中计算出的噪声与测量到的噪声之间的平均比率。头部检查显示出最佳的相关性(测得的与计算得出的噪声比= 1.01)。在胸腔中,计算出的噪声通常高于实测噪声(比率= 0.74),而在腹部,发现相反的结果(比率= 1.20)。