DETERMINATION OF LiF:Mg,Ti AND LiF:Mg,Cu,P TL EFFICIENCY FOR X-RAYS AND THEIR APPLICATION TO MONTE CARLO SIMULATIONS OF DOSEMETER RESPONSE

机译:LiF:Mg,Ti和LiF:Mg,Cu,P TL的X射线效率测定及其在剂量反应的蒙特卡罗模拟中的应用

获取原文
获取原文并翻译 | 示例

摘要

The simulation of response of a new passive area dosemeter for measuring ambient dose equivalent H~(*)(10) for photons has been performed using the Monte Carlo code MCNP and experimentally determined responses of LiF:Mg,Ti and LiF:Mg,Cu,P thermoluminescent (TL) detectors for hard-filtered X-ray spectra from 20 to 300 keV and for ~(137)Cs and ~(60)Co gamma radiation. Relative TL efficiency for both types of detectors, determined in experiments with bare detectors and similar Monte Carlo simulations, compared favourably with prediction of microdosimetric models for proposed micro-dosimetric target sizes in the range of 20-40 nm. The concluding verification experiment showed small deviations between measured and simulated dosemeter energy response values in the range of a few percent.
机译:已经使用蒙特卡罗代码MCNP进行了新的用于测量光子的环境剂量当量H〜(*)(10)的无源区域剂量计的响应仿真,并通过实验确定了LiF:Mg,Ti和LiF:Mg,Cu的响应,P热发光(TL)检测器,用于20至300 keV的硬过滤X射线光谱以及〜(137)Cs和〜(60)Coγ射线。在裸露的探测器和类似的蒙特卡洛模拟实验中确定的两种类型的探测器的相对TL效率与拟议的20-40 nm范围内的微剂量目标尺寸的微剂量模型预测相比具有优势。最后的验证实验表明,测得的和模拟的剂量计能量响应值之间的偏差很小,只有几个百分点。