EVALUATION OF IMAGE QUALITY AND RADIATION DOSE BY ADAPTIVE STATISTICAL ITERATIVE RECONSTRUCTION TECHNIQUE LEVEL FOR CHEST CT EXAMINATION

机译:用自适应统计迭代重建技术进行胸部CT检查的图像质量和辐射剂量评价。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this research is to determine the adaptive statistical iterative reconstruction (ASIR) level that enables optimal image quality and dose reduction in the chest computed tomography (CT) protocol with ASIR. A chest phantom with 0-50 % ASIR levels was scanned and then noise power spectrum (NPS), signal and noise and the degree of distortion of peak signal-to-noise ratio (PSNR) and the root-mean-square error (RMSE) were measured. In addition, the objectivity of the experiment was measured using the American College of Radiology (ACR) phantom. Moreover, on a qualitative basis, five lesions' resolution, latitude and distortion degree of chest phantom and their compiled statistics were evaluated. The NPS value decreased as the frequency increased. The lowest noise and deviation were at the 20 % ASIR level, mean 126.15 + 22.21. As a result of the degree of distortion, signal-to-noise ratio and PSNR at 20 % ASIR level were at the highest value as 31.0 and 41.52. However, maximum absolute error and RMSE showed the lowest deviation value as 11.2 and 16. In the ACR phantom study, all ASIR levels were within acceptable allowance of guidelines. The 20 % ASIR level performed best in qualitative evaluation at five lesions of chest phantom as resolution score 4.3, latitude 3.47 and the degree of distortion 4.25. The 20 % ASIR level was proved to be the best in all experiments, noise, distortion evaluation using ImageJ and qualitative evaluation of five lesions of a chest phantom. Therefore, optimal images as well as reduce radiation dose would be acquired when 20 % ASIR level in thoracic CT is applied.
机译:这项研究的目的是确定自适应统计迭代重建(ASIR)级别,该级别可在具有ASIR的胸部计算机断层扫描(CT)协议中实现最佳图像质量和减少剂量。扫描了具有0-50%ASIR级别的胸部模型,然后进行了噪声功率谱(NPS),信号和噪声以及峰值信噪比(PSNR)的失真度和均方根误差(RMSE) )进行了测量。此外,使用美国放射学院(ACR)体模测量了实验的客观性。此外,在定性的基础上,评估了五个病变的分辨率,胸部幻像的纬度和变形程度以及它们的汇总统计数据。 NPS值随频率增加而降低。最低的噪声和偏差为20%ASIR,平均126.15 + 22.21。作为失真程度的结果,在20%ASIR级别时,信噪比和PSNR最高,分别为31.0和41.52。但是,最大绝对误差和RMSE显示的最低偏差值为11.2和16。在ACR体模研究中,所有ASIR含量均在指南的可接受范围内。 20%ASIR水平在定性评估中对五个人体模型的病变表现最佳,分别为分辨率评分4.3,纬度3.47和畸变度4.25。在所有实验,噪声,使用ImageJ进行失真评估以及对五个人体模型幻影进行定性评估中,20%的ASIR水平被证明是最佳的。因此,当在胸部CT中应用20%的ASIR水平时,将获得最佳图像并降低辐射剂量。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2013年第2期|163-171|共9页
  • 作者单位

    Department of Radiology, Asan Medical Center, Seoul 138-736, Korea,Department of Radiological Science, Eulji University, Seongnam 461-713, Korea;

    Department of Radiology, Kyung Hee University Hospital at Gang-dong, Seoul 134-727, Korea;

    Department of Radiology, Konkuk University Medical Center, Seoul 143-729, Korea;

    Department of Radiologic Technology, Daegu Health College, Daegu 702-722, Korea;

    Department of Radiological Science, Jeonju University, Jeonju 560-759, Korea;

    Department of Radiologic Science, Shinheung University, Uijeongbu 480-701, Korea;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号