首页> 外文期刊>Radiation Protection Dosimetry >APPLICATION OF WHOLE-BODY PERSONAL TL DOSEMETERS IN MIXED FIELD BETA-GAMMA RADIATION
【24h】

APPLICATION OF WHOLE-BODY PERSONAL TL DOSEMETERS IN MIXED FIELD BETA-GAMMA RADIATION

机译:全身性TL剂量计在混合场β-γ辐射中的应用

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Application of whole-body personal TL dosemeters based on a high-sensitivity LiF:Mg,Cu,P (MCP-N) in mixed field beta-gamma radiation has been characterised. The measurements were carried out with ~(90)Sr/~(90)Y, ~(85)Kr and ~(137)Cs point sources to calculate the energy response and linearity of the TLD response in a dose range of 0.1-30 mSv. From the result, calibration curves were obtained, enabling the readout of individual dose equivalent H_p(10) from gamma radiation and H_p(0.07) from beta radiation in mixed field beta-gamma. Limitation of the methodology and its application are presented and discussed.
机译:已经表征了基于高灵敏度LiF:Mg,Cu,P(MCP-N)的全身个人TL剂量计在混合场β-γ辐射中的应用。使用〜(90)Sr /〜(90)Y,〜(85)Kr和〜(137)Cs点源进行测量,以计算剂量范围为0.1-30的TLD响应的能量响应和线性mSv。从结果中获得校准曲线,从而能够在混合场β-γ中从伽马射线中读出单个剂量当量H_p(10),从β射线中读出H_p(0.07)。提出并讨论了该方法的局限性及其应用。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2014年第2期|20-23|共4页
  • 作者单位

    Central Laboratory for Radiological Protection (CLOR), Konwaliowa St. 7, Warsaw 03-194, Poland;

    Central Laboratory for Radiological Protection (CLOR), Konwaliowa St. 7, Warsaw 03-194, Poland,Faculty of Physics, Warsaw University of Technology, Koszykowa St. 75, Warsaw 00-662, Poland;

    Central Laboratory for Radiological Protection (CLOR), Konwaliowa St. 7, Warsaw 03-194, Poland;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词