Current digital mammography (DM) X-ray systems are equipped with advanced automatic exposure control (AEC) systems, which determine the exposure factors depending on breast composition. In the supplement of the European guidelines for quality assurance in breast cancer screening and diagnosis, a phantom-based test is included to evaluate the AEC response to local dense areas in terms of signal-to-noise ratio (SNR). This study evaluates the proposed test in terms of SNR and dose for four DM systems. The glandular fraction represented by the local dense area was assessed by analytic calculations. It was found that the proposed test simulates adipose to fully glandular breast compositions in attenuation. The doses associated with the phantoms were found to match well with the patient dose distribution. In conclusion, after some small adaptations, the test is valuable for the assessment of the AEC performance in terms of both SNR and dose.
机译:当前的数字乳房X线摄影(DM)X射线系统配备有先进的自动暴露控制(AEC)系统,该系统根据乳房成分确定暴露因素。在欧洲乳腺癌筛查和诊断质量保证指南的补充中,包括基于幻像的测试,以信噪比(SNR)评估对局部密集区域的AEC响应。这项研究根据四个DM系统的信噪比和剂量评估了拟议的测试。通过分析计算评估由局部致密区域代表的腺部分。发现所提出的测试模拟了脂肪到完全腺体的乳房成分的衰减。发现与幻影相关的剂量与患者的剂量分布非常匹配。总之,经过一些小改动,该测试对于评估SNR和剂量方面的AEC性能非常有价值。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2016年第4期|143-150|共8页
  • 作者单位

    Dutch Reference Centre for Screening (LRCB), Radboud University Nijmegen Medical Centre (LRCB), PO Box 6873, Nijmegen 6503 GJ, The Netherlands;

    Department of Radiology, UZ Gasthuisberg, Herestraat 49 Box 7003, Leuven 3000, Belgium;

    National Coordinating Centre for Physics in Mammography (NCCPM), Royal Surrey County Hospital, Guildford GU2 7XX,UK,Department of Physics, University of Surrey, Guildford GU2 7XH, UK;

    National Coordinating Centre for Physics in Mammography (NCCPM), Royal Surrey County Hospital, Guildford GU2 7XX,UK,Department of Physics, University of Surrey, Guildford GU2 7XH, UK;

    Dutch Reference Centre for Screening (LRCB), Radboud University Nijmegen Medical Centre (LRCB), PO Box 6873, Nijmegen 6503 GJ, The Netherlands;

    Dutch Reference Centre for Screening (LRCB), Radboud University Nijmegen Medical Centre (LRCB), PO Box 6873, Nijmegen 6503 GJ, The Netherlands;

    Dutch Reference Centre for Screening (LRCB), Radboud University Nijmegen Medical Centre (LRCB), PO Box 6873, Nijmegen 6503 GJ, The Netherlands,Leiden University Medical Centre (LUMC), Albinusdreef 2, Leiden 2333 ZA, The Netherlands;

    Department of Radiology, UZ Gasthuisberg, Herestraat 49 Box 7003, Leuven 3000, Belgium;

    Dutch Reference Centre for Screening (LRCB), Radboud University Nijmegen Medical Centre (LRCB), PO Box 6873, Nijmegen 6503 GJ, The Netherlands;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号