EFFECTIVE DOSE IN TWO DIFFERENT DENTAL CBCT SYSTEMS: NEWTOM VGi AND PLANMECA 3D MID

机译:两种不同的牙科CBCT系统中的有效剂量:NEWTOM VGi和PLANMECA 3D MID

获取原文
获取原文并翻译 | 示例
           

摘要

Cone beam computed tomography, CBCT, is a kind of CT scanner producing conical diverging X-rays, in which a large area of a two-dimensional detector is irradiated in each rotation. Different investigations have been performed on dosimetry of dental CBCT. As there is no special protocol for dental CBCT, CT scan protocols are used for dosimetry. The purpose of this study is measurement of dose to head and neck organs in two CBCT systems, i.e. Planmeca 3D Mid (PM) and NewTom VGi (NT), using thermoluminescence dosimetry and Rando phantom. The thermoluminescent dosimetry (TLD)-100 chips were put at the position of different organs of the head and neck. Two TLD-100 chips were inserted at each position, the dose values were measured for several different field sizes i.e. 8 × 8,12 × 8 and 15 × 15 cm for NewTom, and 10 × 10 and 20 × 17 cm~2 for Planmeca systems. According to the results, the average effective dose in PM is much more than the NT system in the same field size, because of the greater mAs values. For routine imaging protocols used for NT, the effective dose values are 70, 73 and 121 µSv for 8 × 8, 12 × 8 and 15 × 15 cm~2 field sizes, respectively. In PM, the effective dose in 10 × 10 cm~2 and 17 × 20 cm~2 is 259 and 341 µSv, respectively.
机译:锥形束计算机断层扫描技术(CBCT)是一种产生锥形发散X射线的CT扫描仪,每次旋转都会照射较大面积的二维检测器。关于牙科CBCT的剂量测定已进行了不同的研究。由于没有针对牙科CBCT的特殊协议,因此CT扫描协议用于剂量测定。这项研究的目的是使用热致发光剂量法和Rando体模测量两个CBCT系统(即Planmeca 3D Mid(PM)和NewTom VGi(NT))中头颈部器官的剂量。将热发光剂量法(TLD)-100芯片放置在头颈部的不同器官的位置。在每个位置插入两个TLD-100芯片,针对几个不同的场大小测量剂量值,例如NewTom为8×8,12×8和15×15 cm,Planmeca为10×10和20×17 cm〜2系统。根据结果​​,由于更大的mAs值,在相同的场大小下,PM中的平均有效剂量比NT系统大得多。对于用于NT的常规成像方案,对于8×8、12×8和15×15 cm〜2场大小,有效剂量分别为70、73和121 µSv。在PM中,10×10 cm〜2和17×20 cm〜2的有效剂量分别为259和341 µSv。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2017年第3期|287-293|共7页
  • 作者单位

    Nuclear Engineering Department, School of Mechanical Engineering, Shiraz University, Shiraz, Iran;

    Nuclear Engineering Department, School of Mechanical Engineering, Shiraz University, Shiraz, Iran;

    Radiation Sciences Department, Faculty of Allied Medicine, Iran University of Medical Sciences, Tehran, Iran,Radiation Biology Research Center, Iran University of Medical Sciences, Tehran, Iran;

    Nuclear Engineering Department, School of Mechanical Engineering, Shiraz University, Shiraz, Iran,Radiation Research Center, Shiraz University, Shiraz, Iran;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号