The use of pulsed photon radiation in medical, industrial and security sectors has vastly increased during the recent years. The length of pulse from different X-ray flash generators that are commonly used as either portable, battery-operated or fixed systems can be as low as a few femtoseconds. However, the majority of radiation protection instruments, especially various active electronic dosemeters, have limitations when operated in pulsed fields. This study presents measurements that were performed using the dose-integration mode of the RI-02 and Ram Ion survey metres, which are based on ventilated ionisation chamber, when exposed to different pulsed X-ray sources. An intercomparison between the results that were obtained with the survey metres and those that were obtained with thermoluminescence dosemeter, reference passive dosemeter, show good agreement (deviation lies within 10%).
机译:近年来,脉冲光子辐射在医疗,工业和安全领域的使用已大大增加。来自通常用作便携式,电池供电或固定系统的不同X射线闪光灯发生器的脉冲长度可以低至几飞秒。但是,大多数辐射防护仪器,尤其是各种有源电子剂量计,在脉冲场中操作时都有局限性。这项研究提出了在暴露于不同脉冲X射线源的情况下,使用基于通风离子化室的RI-02和Ram离子测量仪的剂量积分模式进行的测量。用测量仪获得的结果与用热致发光剂量仪(参考无源剂量仪)获得的结果之间的比较显示出很好的一致性(偏差在10%以内)。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2017年第3期|297-301|共5页
  • 作者

    D. Ginzburg;

  • 作者单位

    Radiation Detection Department, Rotem Industries Ltd, Rotem Industrial Park 86800, Israel;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);