首页> 外文期刊>Radiation Protection Dosimetry >CHARACTERISATION OF THE PTB THERMAL NEUTRON CALIBRATION FACILITY WITH THE BONNER SPHERE SPECTROMETER NEMUS
【24h】

CHARACTERISATION OF THE PTB THERMAL NEUTRON CALIBRATION FACILITY WITH THE BONNER SPHERE SPECTROMETER NEMUS

机译:用邦纳球光谱仪神经元表征PTB热中子校准装置

获取原文
获取原文并翻译 | 示例
           

摘要

The neutron field of the PTB Thermal Neutron Calibration Facility was characterised with the help of the Bonner sphere spectrometer of PTB, NEutron MUltisphere Spectrometer (NEMUS). For the analysis of the Bonner sphere data an analytical model of the neutron energy distribution was used. The unfolding of neutron spectra was performed via the method of Bayesian parameter estimation. Comparison of the results of thermal neutron fluence and thermal ambient dose equivalent with the results of the reference method of gold activation showed a very good agreement. For fast neutrons the results obtained with NEMUS will serve as reference values for this facility. The facility is ready for use as a thermal neutron reference field for calibrations of neutron measuring devices.
机译:PTB热中子校准设施的中子场借助于PTB的Bonner球光谱仪,NEutron多元光谱仪(NEMUS)进行了表征。为了分析邦纳球数据,使用了中子能量分布的分析模型。中子谱的展开通过贝叶斯参数估计的方法进行。热中子注量和热环境剂量当量的结果与金活化参考方法的结果进行比较显示出很好的一致性。对于快中子,使用NEMUS获得的结果将作为该设施的参考值。该设备已准备好用作热中子参考场,用于中子测量设备的校准。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2018年第4期|98-101|共4页
  • 作者单位

    PTB, Bundesallee 100, D-38116 Braunschweig, Germany;

    PTB, Bundesallee 100, D-38116 Braunschweig, Germany;

    PTB, Bundesallee 100, D-38116 Braunschweig, Germany;

    PTB, Bundesallee 100, D-38116 Braunschweig, Germany;

    PTB, Bundesallee 100, D-38116 Braunschweig, Germany;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);